Troubleshooting and Alarms
71
Check Sensor
Error Message: C54/CHECK
Cause: The Monitor picked up a Sensor current that is outside the operating range or is unstable.
Action:
1.
Press the SEL button and then the ACT button to clear the alarm. The alarm will turn off, and the
HOME screen will reappear. The alarm will repeat periodically until the condition is corrected.
2.
Make sure the insertion site appears normal and the Sensor is still connected. DO NOT REMOVE
CONNECTOR.
3.
If the alarm is triggered two or more times, take out the Sensor. Before putting in a new Sensor, do
the Test Plug procedure at the end of this chapter to check system function.
4.
If the Transmitter is not working correctly, a new Transmitter must be used. If it is working
correctly, put in a new Sensor. Run initialization and calibration. See Chapter 2: Getting Started.
check
12:00
AM
C54
GLUCOSE
SENSOR