D
‐
0125127
‐
C
35
Recommended
separation
distances
between
Portable
and
mobile
RF
communications
equipment
and
the
Avant
REM
The
Avant
REM
is
intended
to
use
in
an
electromagnetic
environment
in
which
radiated
RF
disturbances
are
controlled.
The
customer
or
the
user
of
the
Avant
REM
can
help
prevent
electromagnetic
interference
by
maintaining
a
minimum
distance
between
portable
and
mobile
RF
communications
equipment
(transmitters)
and
the
Avant
REM
as
recommended
below,
according
to
the
maximum
output
power
of
the
communications
equipment.
Rated
maximum
output
power
of
transmitter
W
Separation
distance
according
to
frequency
of
transmitter
meters
150
kHz
to
80
MHz
d
=
1,17
P
80
MHz
to
800
MHz
d
=
1,17
P
800
MHz
to
2,5
GHz
d
=
2,33
P
0,01
0,12
0,12
0,233
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,7
3,7
7,40
100
11,7
11,7
23,3
For
transmitters
rated
at
a
maximum
output
power
not
listed
above,
the
recommended
separation
distance
d
in
meters
(m)
can
be
estimated
using
the
equation
applicable
to
the
frequency
of
the
transmitter,
where
P
is
the
maximum
output
power
rating
of
the
transmitter
in
watts
(W)
according
to
the
transmitter
manufacturer.
NOTE
1
At
80
MHz
and
800
MHz
the
separation
distance
for
the
higher
frequency
range
applies.
NOTE
2
These
guidelines
may
not
apply
in
all
situations.
Electromagnetic
propagation
is
affected
by
absorption
and
reflection
from
structures,
objects
and
people.
A
medical
grade
computer
is
recommended,
conforming
to
the
requirements
of
IEC
60950
‐
1
and
IEC
60601
‐
1
‐
4.
Содержание Avant REM+