Assembly 

 

 

6. Cradle/Swivel 

Bar 

Attachment 

 

a.  Remove nut and bolt from cradle 

 

b. Hold cradle in position at the end of the boom 

and align the holes of the pin and cap with the 
holes in the boom end fork 

 

c. Insert supplied bolt and washer through boom 

fork, cradle cap and pin. Washers must be 
placed on the outside of the fork under bolt 
head and under the nut (Figure 5a, 5b & 5c). 
Place nut on bolt, and secure tightly (Figure 6). 

 

d.  Swivel cradle to check for free movement 

 
7. 

Attach adjusting lever handle into base 

 
8. 

Wheel locks are provided on two casters for 
parking. To lock casters, step on cam lever on 
side of caster. To unlock, step on off lever. 

 

-Warning- 

Never lock wheels or block wheels when lifting 
someone.  Wheel locks are provided for parking 
lifter only  

 

-Warning- 

Inside center attachment points on cradle are 
used for slings, chains or straps.  The four 
outside attachments points are used for 4-point 
slings. 

 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 

Figure 5a 

Figure 5b 

Figure 5c 

Figure 6 

Содержание Patient Lift

Страница 1: ...Manual Hydraulic Patient Lift ...

Страница 2: ...rvice and repair remember your authorized Medline Dealer is able to provide the assistance needed Save this manual for future reference Warning Using other manufacturers parts on the Medline Industries Inc products is unsafe and may result in serious injury to patient and or attendant Use only Medline Industries Inc parts Medline Industries Inc parts are NOT interchangeable with other manufacturer...

Страница 3: ...ain boom points in the same direction as base legs Thread mast safety lock and stud into threaded hole in mast sleeve Tighten firmly locking mast into base Figure 2 Keep mast and base locked at all times except when removing mast from base for storage Warning Be sure the bolt is aligned correctly and fully engaged or failure may result Worn parts could fail resulting in injury to resident caregive...

Страница 4: ...c Place nut on bolt and secure tightly Figure 6 d Swivel cradle to check for free movement 7 Attach adjusting lever handle into base 8 Wheel locks are provided on two casters for parking To lock casters step on cam lever on side of caster To unlock step on off lever Warning Never lock wheels or block wheels when lifting someone Wheel locks are provided for parking lifter only Warning Inside center...

Страница 5: ...draulic pressure knob by turning it counter clockwise not more than a quarter turn Figure 7 2 Guide descent of the patient Push gently on the patient s knees to attain correct sitting posture while lowering 3 After patient reaches seat push down on boom to slacken straps or chains Warning Lifters can tip over Keep base widened for stability Never lock wheel locks or block wheels when lifting someo...

Страница 6: ...etween the base legs X X 5 Cradle Check hanger post for wear or damage X Check sling attachments for wear X X 6 Swivel Bar Check nut at bottom of swivel bar hook and eye hook on boom wear X Check for 2 plastic washers between nut of swivel bar hook and underside of swivel bar assembly X X 7 Pump Assembly Pump handle and valve work smoothly X X Check for oil leakage a light film is normal but not t...

Страница 7: ...on suspended in a sling over ANY distance is NOT recommended The lifter is NOT designed for use as an equipment hoist NEVER use the lifter to lift or transfer anything other than a patient This can result in damage to the lift Transferring DO NOT attempt any transfer without approval of the patient s physician nurse or medical attendant Thoroughly read the instructions in the owner s manual observ...

Страница 8: ...decrease the distance you have to elevate the patient B To place the sling under the patient the instructions are as follows 1 Place the folded sling behind the patient s back Carefully roll the patient onto his side 2 Pull the leg loops forward and under patient s thighs 3 Cross the loops Warning Pulling sharply on sling may cause the patient to fall forward resulting in injury Always stand in fr...

Страница 9: ...ent onto the bed E Remove the chains or straps from the sling Transferring Patients in Lifter A During transfer ensure that the patient is facing the attendant and keep the patient as low as possible so that the patient s feet rest on the base of the lifter straddling the mast A lower center of gravity reduces the risk of a tip over Warning DO NOT roll the caster base over objects such as shag or ...

Страница 10: ...raps attached to sling Removing Sling from Under Patient Note It is not necessary to remove the sling unless desired by the patient A Remove the armrests from the chair if possible optional Disconnect the sling from the lifter B Stand in front of the patient lean the patient forward and place his arms on his knees for stability Pull the sling flaps to the side out from under patient s thighs C Sti...