Medline MDS400SA Скачать руководство пользователя страница 9

MDS400&600SA (GPI252) 

Revision: 02/16/16

8

MAINTENANcE & INsPEcTION chEckINg LIsT

The operator of the lift is to inspect the Medline MDS400EL before each use. Check all bolts for tightness. Make sure the 
base can be easily opened, and that all lift parts are in place. Make sure all casters turn freely, and that the caster brakes 
can be engaged. Make certain all necessary items (i.e. slings, wheel chairs, etc.) are accessible and ready for use. See the 
instructions below to check the battery capacity and appropriate re-charge levels.

At least once a month, the lift should be thoroughly inspected by a qualified technician to recognize any signs of wear, 
and/or looseness of bolts or parts. Replace any worn parts immediately.

To lubricate, place a drop of general purpose oil on the following points when the Medline MDS400SA or MDS600SA is 
serviced and every month thereafter:

• 

Top of mast bolt connection

• 

Spreader bar hinge

• 

Caster axles

chEckINg BATTERY cAPAcITY

Check batteries by pressing on the control box LCD screen or by pressing the battery indicator on the battery pack.

• 

Make sure the battery box red reset/emergency stop button is “ON” (i.e. the red button is up).

• 

Insert the AC charging cable into the control box and pug into a working electrical outlet.

• 

Batteries require 7-9 hours to be fully charged.

• 

A complete initial charge should take approximately 12 hours. Subsequent recharges should take 4 hours or less. Do 
not charge the batteries over 12 hours to avoid premature charger wear. Charging longer than 12 hours will not affect 
the battery life.

• 

Unplug the AC charger cable FIRST before using the lift.

• 

Replace the two internal batteries when frequent charging is required.

• 

*The Medline MDS400SA & MDS600SA will not operate while the batteries are being charged.

Содержание MDS400SA

Страница 1: ...BATTERY OPERATED PATIENT LIFT OWNER S MANUAL MDS400SA MDS600SA Not made with natural rubber latex...

Страница 2: ...ping damage Refuse delivery if the package appears to be badly damaged If the merchandise is received damaged or is missing components contact the shipping company immediately and file a claim For fur...

Страница 3: ...t that can be safely lifted is 400 lbs for the MDS400SA and 600 lbs for the MDS600SA NEVER EXCEED SAFE WORKING LOAD CAPACITY The Medline MDS400SA or MDS600SA allows a resident to be lifted and transfe...

Страница 4: ...of non Medline Lift parts and components WARNING Avoid personal injury by making sure the electrical connector is disconnected before conducting any electrical inspection Contact your dealer if any e...

Страница 5: ...for 12 hours 2 Attach boom to top of mast Re insert the bolts into the holes at top of the mast and tighten the nuts 3 1 Attach the bottom actuator to the bracket on the mast and insert pin Insert kee...

Страница 6: ...life Battery capacity indicator on battery pack Soft start and stop for lifting actuator Soft and Hardware over current protection Actuator s Over duty protection with sign on LCD Emergency stop butt...

Страница 7: ...ED THAT THE BUTTON IS PRESSED DOWN WHEN THE LIFT WILL NOT BE USED FOR OVER 3 DAYS TO MAINTAIN OPTIMAL BATTERY LIFE AND PERFORMANCE PATIENT LIFT WILL NOT OPERATE WHILE CHARGING KG KG Low battery Capaci...

Страница 8: ...prevent water ingress REMOVE AND INSERT BATTERY PACK REMOVE Press button A and pull out in the up direction Fig 1 INSERT Insert the battery pack into control box as show in Fig 2 3 While pressing the...

Страница 9: ...g points when the Medline MDS400SA or MDS600SA is serviced and every month thereafter Top of mast bolt connection Spreader bar hinge Caster axles CHECKING BATTERY CAPACITY Check batteries by pressing...

Страница 10: ...easily on its casters and the casters are firmly attached to base If lift won t roll freely check caster surface for wear and replace if required Lubricate caster axles at each maintenance session El...

Страница 11: ...l All controls are easy to use and do not require excessive force DO NOT park a loaded lift on ANY sloping surface DO NOT use electric lifts in a shower DO NOT charge an electric lift in a bathroom or...

Страница 12: ...Position sling behind the back with the two red pads under the arms and the single red lumbar pad against the small of the back 3 Make sure the adjustment strap is on the outside 4 Cross the chest be...

Страница 13: ...the kneepad Attach the sling straps to the hooks Have the patient s hands holding the handles For patients who cannot hold the handles have them hold their arms around the chest Press the UP button on...

Страница 14: ...nually until the patient has been lowered NOTE THE EMERGENCY LOWERING DEVICE IS INTENDED FOR USE DURING LIFT FAILURE THIS DEVICE WILL ALLOW LOWERING OF PATIENTS ONLY LIFT SPECIFICATION CHART ITEM NUMB...

Страница 15: ...charge Control Box Charger still plugged in Poor connection between control box and hand control or defective hand control Poor connection between actuator and control box Unplug charger Plug hand con...

Страница 16: ...odel Name Number and Serial Number on the packaging 2 Proof of purchase i e copy of the original invoice from either Medline or its dealer 3 A Return Authorization number RA obtained by calling Medlin...

Страница 17: ...medline com Medline United States 1 800 MEDLINE 633 5463 www medline com info medline com Medline Industries Inc One Medline Place Mundelein IL 60060 Some products may not be available for sale in Mex...

Отзывы: