MDS400&600SA (GPI252) 

Revision: 02/16/16

14

TROuBLEshOOTINg chART

sYMPTOMs

cOMPONENTs

cONcERNED

POssIBLE cAusEs

AcTION

sOLuTION

The actuator is 

installed and a clicking 

sound is audible.

-Actuator

-Actuator and 

PCB

-Actuator plug is not sealed 

properly into control box.

-Actuator is defective or

completed its life cycle.

-Control box is defective.

Test control box 

on another

Medline lift.

-Firmly plug the actuator plug 

into control box.

-Replace actuator if control box 

is good.

-Replace PCB if control box is 

bad.

The actuator will 

either go up or down, 

but not both.

Hand Control 

or PCB

Hand control or control box 

defective.

Test hand 

control on 

another

Medline lift.

-Replace PCB if hand control is 

good.

-Replace hand control if it is bad.

The actuator makes a 

grinding sound while 

lifting, but works fine 

without a load.

Actuator

The internal gear is stripped 

likely due to overloading

Stop using the 

lift.

Repair or replace the actuator.

The actuator stops 

and starts during 

lifting or lowering.

Batteries

Bad batteries

Check battery 

in-dicator. 

Confirm if lights 

drop quickly to 

red light.

Replace batteries.

The lift will not 

operate even though 

battery indicator 

shows charge.

Control Box

-Charger still plugged in.

-Poor connection between 

control box and hand control, 

or defective hand control.

-Poor connection between 

actuator and control box.

-Unplug charger.

-Plug hand control firmly into 

control box, or replace hand 

control.

-Plug actuator cord firmly into 

control box.

The lift will not 

operate and battery 

indicator has no lights 

illuminated.

Control box 

and battery

-Control box may be in “off” 

position.

-Internal batteries may be 

loose from their connections.

-Batteries are in need of being 

charged.

-Batteries are dead and won’t 

hold a new charge.

-Turn the red emergency stop 

button until it pops up to the “on” 

position.

-Check inside control box to 

ensure batteries are connected.

-Use the charger to charge the 

batteries making sure the red 

emergency stop is in “on” or “up” 

position.

-Replace the two internal 

batteries.

During charging the 

battery, indicator is 

not illuminated and 

won’t charge.

-Control Box

-Charger

-The red emergency stop 

button is in “off” or “down” 

position.

-The charger is defective or at 

end of it’s life cycle.

-Turn the red emergency stop 

button until it pops up into the 

“on” position.

-Inspect the charger for damage 

and replace if required.

The battery indicator 

indicates need for 

frequent charging 

and/or lift looses 

power.

Batteries

The internal batteries are at 

the end of life cycle.

Replace the two internal 

batteries.

Содержание MDS400SA

Страница 1: ...BATTERY OPERATED PATIENT LIFT OWNER S MANUAL MDS400SA MDS600SA Not made with natural rubber latex...

Страница 2: ...ping damage Refuse delivery if the package appears to be badly damaged If the merchandise is received damaged or is missing components contact the shipping company immediately and file a claim For fur...

Страница 3: ...t that can be safely lifted is 400 lbs for the MDS400SA and 600 lbs for the MDS600SA NEVER EXCEED SAFE WORKING LOAD CAPACITY The Medline MDS400SA or MDS600SA allows a resident to be lifted and transfe...

Страница 4: ...of non Medline Lift parts and components WARNING Avoid personal injury by making sure the electrical connector is disconnected before conducting any electrical inspection Contact your dealer if any e...

Страница 5: ...for 12 hours 2 Attach boom to top of mast Re insert the bolts into the holes at top of the mast and tighten the nuts 3 1 Attach the bottom actuator to the bracket on the mast and insert pin Insert kee...

Страница 6: ...life Battery capacity indicator on battery pack Soft start and stop for lifting actuator Soft and Hardware over current protection Actuator s Over duty protection with sign on LCD Emergency stop butt...

Страница 7: ...ED THAT THE BUTTON IS PRESSED DOWN WHEN THE LIFT WILL NOT BE USED FOR OVER 3 DAYS TO MAINTAIN OPTIMAL BATTERY LIFE AND PERFORMANCE PATIENT LIFT WILL NOT OPERATE WHILE CHARGING KG KG Low battery Capaci...

Страница 8: ...prevent water ingress REMOVE AND INSERT BATTERY PACK REMOVE Press button A and pull out in the up direction Fig 1 INSERT Insert the battery pack into control box as show in Fig 2 3 While pressing the...

Страница 9: ...g points when the Medline MDS400SA or MDS600SA is serviced and every month thereafter Top of mast bolt connection Spreader bar hinge Caster axles CHECKING BATTERY CAPACITY Check batteries by pressing...

Страница 10: ...easily on its casters and the casters are firmly attached to base If lift won t roll freely check caster surface for wear and replace if required Lubricate caster axles at each maintenance session El...

Страница 11: ...l All controls are easy to use and do not require excessive force DO NOT park a loaded lift on ANY sloping surface DO NOT use electric lifts in a shower DO NOT charge an electric lift in a bathroom or...

Страница 12: ...Position sling behind the back with the two red pads under the arms and the single red lumbar pad against the small of the back 3 Make sure the adjustment strap is on the outside 4 Cross the chest be...

Страница 13: ...the kneepad Attach the sling straps to the hooks Have the patient s hands holding the handles For patients who cannot hold the handles have them hold their arms around the chest Press the UP button on...

Страница 14: ...nually until the patient has been lowered NOTE THE EMERGENCY LOWERING DEVICE IS INTENDED FOR USE DURING LIFT FAILURE THIS DEVICE WILL ALLOW LOWERING OF PATIENTS ONLY LIFT SPECIFICATION CHART ITEM NUMB...

Страница 15: ...charge Control Box Charger still plugged in Poor connection between control box and hand control or defective hand control Poor connection between actuator and control box Unplug charger Plug hand con...

Страница 16: ...odel Name Number and Serial Number on the packaging 2 Proof of purchase i e copy of the original invoice from either Medline or its dealer 3 A Return Authorization number RA obtained by calling Medlin...

Страница 17: ...medline com Medline United States 1 800 MEDLINE 633 5463 www medline com info medline com Medline Industries Inc One Medline Place Mundelein IL 60060 Some products may not be available for sale in Mex...

Отзывы: