Medline MDS400SA Скачать руководство пользователя страница 11

MDS400&600SA (GPI252) 

Revision: 02/16/16

10

DO NOT OPERATE LIFT uNLEss
ALL MAINTENANcE POINTs PAss INsPEcTION

WARNINg:

•  ALWAYS

 plan your lifting operations before commencing.

•  ALWAYS

 carry out the daily checklist before using the lift.

•  ALWAYS

 familiarize yourself with the operating control and safety features of a lift before lifting a patient.

•  DO NOT

 use a sling unless it is recommended for use with the lift.

•  ALWAYS

 check if the sling is suitable for the particular patient and is of the correct size and capacity.

•  NEVER

 use a sling which is frayed or damaged.

•  ALWAYS

 fit the sling according to the instructions provided.

•  ALWAYS

 check the safe working load of the lift to be sure it is suitable for the weight of the patient.

•  ALWAYS

 carry out lifting operations according to the instructions in the user manual.

•  NEVER

 disconnect or bypass a control or safety feature because it seems easier to operate the lift.

•  DO NOT

 lift a patent with the caster brakes on. Always let the lift find the correct center of gravity.

•  DO NOT

 attempt to maneuver the lift by pushing on the mast, boom, or patient.

•  ALWAYS

 maneuver the lift with the handle provided.

•  ALWAYS

 lower the patient to the lowest comfortable position before transfers.

•  DO NOT

 push a loaded lift at speeds which exceed a slow walking pace (1.9 miles/hour, 2.6 foot/second).

•  DO NOT

 push the lift over uneven or rough ground. Particularly if loaded.

•  DO NOT 

attempt to push/pull a loaded lift over a floor obstruction, which the casters are unable to ride over.

•  NEVER

 force an operating/safety control. All controls are easy to use and do not require excessive force.

•  DO NOT

 park a loaded lift on 

ANY

 sloping surface.

•  DO NOT

 use electric lifts in a shower.

•  DO NOT

 charge an electric lift in a bathroom or shower room.

•  DO NOT 

lift a patient unless you are trained and competent to do so.

• 

Your lift is for patient lifting. 

DO NOT

 use it, or allow it to be used for any other purpose.

•  DO NOT

 bump the lift down steps, loaded, or unloaded.

•  DO NOT

 attempt to negotiate a loaded lift on a slope.

•  ALWAYS

 check that the hoist is not charging before moving as the electrical connection may be damaged.

•  CAUTION

: Keep the batteries fully charged. Place the lift on charge whenever it is not in use. The charger will not 

allow the batteries to “overcharge”.

•  NEVER

 lift with the legs in the closed/transport position. The closed position is for storage and transport only.

•  CAUTION

: Never run the batteries completely flat. If the audible warning sounds, complete the lifting operation in 

hand and place the lift on charge.

•  CAUTION

: Never store the lift for long periods without regular charging throughout the storage period.

•  CAUTION

: Never disconnect the charger plug by pulling on the cable.

•  T

he lift has two braked casters, which can be applied for parking. When lifting, the casters should be left free and 

LOCKED

. Keep casters 

UNLOCKED

 when lowering.

•  DO NOT

 apply the brakes because if the brakes are applied, the patient might swing to the center of gravity and this 

may prove disconcerting and uncomfortable to the resident.

•  NOTE

: Lock casters when raising the patient; keep them unlocked while lowering the patient. The reason for locking 

while raising is the patient could potentially kick the unit away or move it away, so the knee brace will not be lined up 
properly.

Содержание MDS400SA

Страница 1: ...BATTERY OPERATED PATIENT LIFT OWNER S MANUAL MDS400SA MDS600SA Not made with natural rubber latex...

Страница 2: ...ping damage Refuse delivery if the package appears to be badly damaged If the merchandise is received damaged or is missing components contact the shipping company immediately and file a claim For fur...

Страница 3: ...t that can be safely lifted is 400 lbs for the MDS400SA and 600 lbs for the MDS600SA NEVER EXCEED SAFE WORKING LOAD CAPACITY The Medline MDS400SA or MDS600SA allows a resident to be lifted and transfe...

Страница 4: ...of non Medline Lift parts and components WARNING Avoid personal injury by making sure the electrical connector is disconnected before conducting any electrical inspection Contact your dealer if any e...

Страница 5: ...for 12 hours 2 Attach boom to top of mast Re insert the bolts into the holes at top of the mast and tighten the nuts 3 1 Attach the bottom actuator to the bracket on the mast and insert pin Insert kee...

Страница 6: ...life Battery capacity indicator on battery pack Soft start and stop for lifting actuator Soft and Hardware over current protection Actuator s Over duty protection with sign on LCD Emergency stop butt...

Страница 7: ...ED THAT THE BUTTON IS PRESSED DOWN WHEN THE LIFT WILL NOT BE USED FOR OVER 3 DAYS TO MAINTAIN OPTIMAL BATTERY LIFE AND PERFORMANCE PATIENT LIFT WILL NOT OPERATE WHILE CHARGING KG KG Low battery Capaci...

Страница 8: ...prevent water ingress REMOVE AND INSERT BATTERY PACK REMOVE Press button A and pull out in the up direction Fig 1 INSERT Insert the battery pack into control box as show in Fig 2 3 While pressing the...

Страница 9: ...g points when the Medline MDS400SA or MDS600SA is serviced and every month thereafter Top of mast bolt connection Spreader bar hinge Caster axles CHECKING BATTERY CAPACITY Check batteries by pressing...

Страница 10: ...easily on its casters and the casters are firmly attached to base If lift won t roll freely check caster surface for wear and replace if required Lubricate caster axles at each maintenance session El...

Страница 11: ...l All controls are easy to use and do not require excessive force DO NOT park a loaded lift on ANY sloping surface DO NOT use electric lifts in a shower DO NOT charge an electric lift in a bathroom or...

Страница 12: ...Position sling behind the back with the two red pads under the arms and the single red lumbar pad against the small of the back 3 Make sure the adjustment strap is on the outside 4 Cross the chest be...

Страница 13: ...the kneepad Attach the sling straps to the hooks Have the patient s hands holding the handles For patients who cannot hold the handles have them hold their arms around the chest Press the UP button on...

Страница 14: ...nually until the patient has been lowered NOTE THE EMERGENCY LOWERING DEVICE IS INTENDED FOR USE DURING LIFT FAILURE THIS DEVICE WILL ALLOW LOWERING OF PATIENTS ONLY LIFT SPECIFICATION CHART ITEM NUMB...

Страница 15: ...charge Control Box Charger still plugged in Poor connection between control box and hand control or defective hand control Poor connection between actuator and control box Unplug charger Plug hand con...

Страница 16: ...odel Name Number and Serial Number on the packaging 2 Proof of purchase i e copy of the original invoice from either Medline or its dealer 3 A Return Authorization number RA obtained by calling Medlin...

Страница 17: ...medline com Medline United States 1 800 MEDLINE 633 5463 www medline com info medline com Medline Industries Inc One Medline Place Mundelein IL 60060 Some products may not be available for sale in Mex...

Отзывы: