PULSE OXIMETER
- 20 -
HCSMD400
6.3 Calibrating
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor
or sensor. Clinical testing is used to establish the SpO
2
accuracy. The measured arterial
hemoglobin saturation value (SpO
2
) of the sensors is compared to arterial hemoglobin
oxygen (SaO
2
) value, determined from blood samples with a laboratory CO-oximeter. The
accuracy of the sensors in comparison to the CO-oximeter samples measured over the SpO
2
range of 70%
~
99%. Accuracy data is calculated using the root-mean-squared (Arms value)
for all subjects, per ISO 9919:2005, Medical Electrical Equipment–Particular requirements for
the basic safety and essential performance of pulse oximeter equipment for medical use.
6.4 Trouble Shooting
a) The oximeter can’t be powered on
Please check the batteries. If you use the external power supply, please check if the power
supply is connected with oximeter properly.
b) “Probe off” alarm
Please check if the probe is connected with the oximeter correctly. If the sensor is with
extension cable please check if the extension cable is connected with the sensor correctly.
c) “No finger” alarm
Please check whether the sensor is correctly connected with patient’s finger.
6.5 Warranty and Repair
6.5.1 Maintenance Method
a) Maintenance responding time: 9:00am~17:30pm, Monday to Friday
b) Service support: our company offers the support by hot line, e-mail or spares parts.
Spare parts: our company changes parts if it is necessary free of charge in the warranty
period.
Because parts are the sources of maintenance, user should send them back to our company
if not specified.
c) Update the system software free of charge.
6.5.2 Exempt and Limitation
a) Our company isn’t responsible for such damage caused by force majeure. For example:
fire, thunder flash, flood, cyclone, hail, earthquake, house collapse, commotion, plane failing
and traffic accident, deliberate damage, lack of fuel or water, labor and capital bother, strike
and stop-working etc.
b) Non-service items
● The corresponding charge and insurance charge of disassembling, refurbishing,
repackaging and moving the oximeter or the part of it.
● The damage caused by the third company not commended by our company to adjust,
install replace the parts of the oximeters.
● The damage and failure caused by the users incorrect operations not complying with the
operator’s manual.
c) Our company will not provide the free maintenance in the warranty if the oximeter is