DO NOT touch the sampling end of the test strip with the bottle of control
solution. See illustration of sampling end below.
Step 5:
Gently touch the tip of the test strip to the drop of control solution. The meter will
beep indicating it has received the control solution. The screen will start to count
down. After 6 seconds, the control solution test result will appear on the meter
display screen.
Step 6:
Compare the reading on the screen to the Low range printed
on the test strip bottle. If the reading falls outside the Low range
printed on the test strip bottle, see Control Solution
Troubleshooting, on page 31.
NOTE:
• Control solution test results will be stored in meter
memory and indicated by ctl icon.
• Control solution test results will not be used for
calculating averages.
Step 7:
Remove the used test strip. The meter will
automatically turn off. Discard the used test
strip. Follow steps 8 through 14 to perform
the High control solution testing.
EvenCareG2
Control Solution Testing
29
53
79
135
203
Low
Low
High
High
IMPORTANT:
Do not reuse test strips.
Содержание evencare g2