Medline DYND70800 Скачать руководство пользователя страница 8

12

13

Recommended Separation Distance Between 

Portable and Mobile RF Communications Equipment 

and Medline Electrical Clipper Charging Base,  

DYND70800/DYND70801/DYND70802 Series

REF DYND70800/DYND70801/DYND70802 Series is intended for use in the 
electromagnetic environment in which RF disturbances are controlled. The 
customer or the user of REF DYND70800/DYND70801/DYND70802 Series can 
help prevent electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment (transmitters) and 
REF DYND70800/DYND70801/DYND70802 Series as recommended below, 
according to the maximum output power of the communications equipment.

Rated Maximum 

Output Power of 

Transmitter W

Separations Distance According to  

Frequency of Transmitter m

150 kHz to  

80 MHz 

d = 1.2 √P

80 kHz to  

800 MHz

d = 1.2 √P

800 kHz to  

2.5 GHz

d = 2.3 √P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the 

recommended separation distance d in meters (m) can be estimated using the 

equation applicable to the frequency of the transmitter, where P is the maximum 

output power rating of the transmitter in watts (W) according to the transmitter 

manufacturer.

NOTE 1:

 at 80 MHz and 800 MHz, the separation distance for the higher 

frequency range applies.

NOTE 2:

 These guidelines may not apply in all situations. Electromagnetic 

propagation is affected by absorption and reflection from structures, objects  

and people.

NOTE 1:

 at 80 MHz and 800 MHz, the separation distance for the higher 

frequency range applies.

NOTE 2: 

These guidelines may not apply in all situations. Electromagnetic 

propagation is affected by absorption and reflection from structures, objects 
and people.

a.

  Field strengths from fixed transmitters, such as base stations for radio 

(cellular/cordless) telephones and land mobile radios, amateur radio, AM 
and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed 
RF transmitters, an electromagnetic site survey should be considered. If 
the measured field strength in the location in which REF DYND70800/
DYND70801/DYND70802 Series is used exceeds the applicable RF compliance 
level above, REF DYND70800/DYND70801/DYND70802 Series should be 
observed to verify normal operation. If abnormal performance is observed, 
additional measures may be necessary, such as reorienting or relocation REF 
DYND70800/DYND70801/DYND70802 Series.

b.

  Over the frequency range 150 kHz to 80 MHz, field strengths should be less 

than 2 V/m.

Содержание DYND70800

Страница 1: ...MEDLINE ELECTRICAL CLIPPER WITH CHARGING BASE INSTRUCTIONS FOR USE DYND70800 DYND70801 DYND70802 Wet Dry Charge Off O...

Страница 2: ...ade attachments are interchangeable to fit onto the Electric Surgical Clipper depending on the use of the attachment INDICATIONS FOR USE The Electric Surgical Clipper is intended for the removal of bo...

Страница 3: ...read and understand all safety information before using 1 Use this clipper only for its intended use as explained in this insert 2 The clipper must only be used with the charger provided 3 Do not sub...

Страница 4: ...en not in use after cleaning and drying place clipper body in the charging adapter for charging and safe storage Contact your Medline Representative to order replacement blades The blades are intended...

Страница 5: ...ne Electrical Clipper Charging Base DYND70800 DYND70801 DYND70802 Series should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary Medline Electrical Clipp...

Страница 6: ...ital environment Guidance and Manufacturer s Declaration Electromagnetic Emission Medline Electrical Clipper Charging Base DYND70800 DYND70801 DYND70802 Series is intended for use in the electromagnet...

Страница 7: ...maximum output power rating of the transmitter in watts W according to the transmit ter manufacture and d is the recommended separation distance in meters m Field strength from fixed RF transmitters a...

Страница 8: ...er in watts W according to the transmitter manufacturer NOTE 1 at 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situati...

Страница 9: ...ale lectrique est destin e au rasage des poils du corps des patients avant une intervention chirurgicale ou d autres op rations mineures La tondeuse peut tre utilis e de fa on s curitaire sur les poil...

Страница 10: ...ts mouvements l gers Pour obtenir de meilleurs r sultats raser dans le sens inverse de la pousse des poils Dans les endroits difficiles bien tendre la peau pendant le rasage MISES EN GARDE Toujours pr...

Страница 11: ...minutes ou une vie utile maximale de 5 ans sur le terrain Inspectez la tondeuse pour d celer tout dommage avant l utilisation et jetez la ou cessez de l utiliser si elle ne fonctionne pas Environneme...

Страница 12: ...re Si cet appareil cause des interf rences nuisibles aux autres dispositifs ce qui peut tre d termin en allumant et en teignant l appareil on encourage l utilisateur essayer de corriger ces interf ren...

Страница 13: ...riations de tension missions de scintille ment CEI 61000 3 3 Conforme missions RF CISPR 11 Environnement d tablisse ment de soins de sant professionnels Comme le groupe 1 Renseignements sur la CEM rel...

Страница 14: ...802 est con u pour tre utilis dans l environnement lectromagn tique sp cifi ci dessous Le client ou l utilisateur du chargeur de la tondeuse lectrique Medline s ries DYND70800 DYND70801 DYND70802 doit...

Страница 15: ...ieures 2 V m Directive et d claration du fabricant immunit lectromagn tique Les s ries DYND70800 DYND70801 DYND70802 sont pr vues pour tre utilis es dans l environnement lectromagn tique pr cis ci des...

Страница 16: ...l quipement de communication Puissance de sortie maximale nominale de l metteur W Distance de s paration selon la fr quence de l metteur m 150 kHz 80 MHz d 1 2 P 80 kHz 800 MHz d 1 2 P 800 kHz 2 5 GHz...

Страница 17: ...www medline com Manufactured for Fabriqu pour Medline Industries Inc Three Lakes Drive Northfield IL 60093 USA 1 800 MEDLINE Made in China Fabriqu en Chine RE21WZA...

Отзывы: