Medline DYND70800 Скачать руководство пользователя страница 4

4

5

Specifications:

Power Source: 

AC 100 – 240V  50-60Hz 0.2A

Charger Output/Clipper Input: 

DC 3.3V 100mA

Motor Voltage: 

DC 2.4V

Mode of Operation: 

Continuous Operation

Battery: 

2x Ni-MH (AA 600mAh  - 1.2 Vdc)

Product Life Use: 

This device has been tested to confirm operational functionality for more 

than 50,000 15-minute uses, or up to 5 years of expected unit use in the 

field. Inspect for damage prior to use and discard/stop use if clipper not 

functioning.

Environment:            

Operation                         

Storage Transport

Temperature:            

+0°C to +40°C

-10°C to +60°C 

Humidity:                  

 30% - 75% RH

10% - 100% RH

Pressure:           

700hPa – 1060hPa

500hPa – 1060hPa

!

 Caution- Read accompanying documentation

  Double Insulated 

Watertight equipment

 Type B Applied Part

DYND70800/DYND70801/DYND70802 SURGICAL CLIPPER 
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL 
HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI STD.
ES60601-1 AND CAN/CSA STD. C22.2 NO.60601-1

Notes:

•  Charge at room temperature of +41°F (+5°C) to +95°F (+35°C).
•  Avoid charge in direct sunlight or near radiating heat.

To remove blade: 

Make sure the On/Off switch is in the off position. Position the clipper over    

an appropriate waste receptacle with the metal part of the blade facing 

downward and the back of the clipper facing up.  Push the blade away 

from the clipper to release the blade. When not in use, after cleaning 

and drying, place clipper body in the charging adapter for charging and 

safe storage. Contact your Medline Representative to order replacement 
blades. The blades are intended for single patient use.

Item numbers:

DYND70880: Universal Blade, Single Use, 50/cs
UBLADEFL: Universal Blade, Flat Style, Single Use, 50/cs
DYND70885: Coarse Hair Blade, Single Use, 50/cs
CBLADEFL: Coarse Hair Blade, Flat Style, Single Use, 50/cs
DYND70890: Narrow Blade, Single Use, 50/cs
NBLADEFL: Narrow Blade, Flat Style, Single Use, 50/cs
CSBLADE:  Close Shave Blade, Single Use, 50/cs 

Wall Mount Instructions: (Optional)

To install optional wall mount:

1.

  Select an appropriate location on the wall using the screws provided.

2.

    Attach the detachable wall mount with screws making sure that the 

arrow on the back (flush part) points up.

3.

   Slide the charging adapter grooves downward into the wall mount 

until it clicks. To remove the charging adapter from the wall mount, 

disengage and slide the adapter upward. 

Care and Cleaning of the Clipper Body: 

Always unplug the charging adapter before cleaning. To remove loose hair 
and other gross debris from clipper, place under running water or soak in a 
water bath. Wipe the clipper with a damp cloth containing a non-corrosive 
disinfectant, such as: Isopropyl alcohol, OPA (Ortho-phthaldehyde), 2% 
glutaraldehyde or any other non-corrosive disinfectant wipe. Dry clipper 
with a clean towel. Return clipper to either the charging adapter or to a 
storage area.

Содержание DYND70800

Страница 1: ...MEDLINE ELECTRICAL CLIPPER WITH CHARGING BASE INSTRUCTIONS FOR USE DYND70800 DYND70801 DYND70802 Wet Dry Charge Off O...

Страница 2: ...ade attachments are interchangeable to fit onto the Electric Surgical Clipper depending on the use of the attachment INDICATIONS FOR USE The Electric Surgical Clipper is intended for the removal of bo...

Страница 3: ...read and understand all safety information before using 1 Use this clipper only for its intended use as explained in this insert 2 The clipper must only be used with the charger provided 3 Do not sub...

Страница 4: ...en not in use after cleaning and drying place clipper body in the charging adapter for charging and safe storage Contact your Medline Representative to order replacement blades The blades are intended...

Страница 5: ...ne Electrical Clipper Charging Base DYND70800 DYND70801 DYND70802 Series should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary Medline Electrical Clipp...

Страница 6: ...ital environment Guidance and Manufacturer s Declaration Electromagnetic Emission Medline Electrical Clipper Charging Base DYND70800 DYND70801 DYND70802 Series is intended for use in the electromagnet...

Страница 7: ...maximum output power rating of the transmitter in watts W according to the transmit ter manufacture and d is the recommended separation distance in meters m Field strength from fixed RF transmitters a...

Страница 8: ...er in watts W according to the transmitter manufacturer NOTE 1 at 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situati...

Страница 9: ...ale lectrique est destin e au rasage des poils du corps des patients avant une intervention chirurgicale ou d autres op rations mineures La tondeuse peut tre utilis e de fa on s curitaire sur les poil...

Страница 10: ...ts mouvements l gers Pour obtenir de meilleurs r sultats raser dans le sens inverse de la pousse des poils Dans les endroits difficiles bien tendre la peau pendant le rasage MISES EN GARDE Toujours pr...

Страница 11: ...minutes ou une vie utile maximale de 5 ans sur le terrain Inspectez la tondeuse pour d celer tout dommage avant l utilisation et jetez la ou cessez de l utiliser si elle ne fonctionne pas Environneme...

Страница 12: ...re Si cet appareil cause des interf rences nuisibles aux autres dispositifs ce qui peut tre d termin en allumant et en teignant l appareil on encourage l utilisateur essayer de corriger ces interf ren...

Страница 13: ...riations de tension missions de scintille ment CEI 61000 3 3 Conforme missions RF CISPR 11 Environnement d tablisse ment de soins de sant professionnels Comme le groupe 1 Renseignements sur la CEM rel...

Страница 14: ...802 est con u pour tre utilis dans l environnement lectromagn tique sp cifi ci dessous Le client ou l utilisateur du chargeur de la tondeuse lectrique Medline s ries DYND70800 DYND70801 DYND70802 doit...

Страница 15: ...ieures 2 V m Directive et d claration du fabricant immunit lectromagn tique Les s ries DYND70800 DYND70801 DYND70802 sont pr vues pour tre utilis es dans l environnement lectromagn tique pr cis ci des...

Страница 16: ...l quipement de communication Puissance de sortie maximale nominale de l metteur W Distance de s paration selon la fr quence de l metteur m 150 kHz 80 MHz d 1 2 P 80 kHz 800 MHz d 1 2 P 800 kHz 2 5 GHz...

Страница 17: ...www medline com Manufactured for Fabriqu pour Medline Industries Inc Three Lakes Drive Northfield IL 60093 USA 1 800 MEDLINE Made in China Fabriqu en Chine RE21WZA...

Отзывы: