![Medline DYND70800 Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/medline/dynd70800/dynd70800_instructions-for-use-manual_1762593006.webp)
8
9
Guidance and Manufacturer’s Declaration
– Electromagnetic Immunity –
Medline Electrical Clipper Charging Base, DYND70800/DYND70801/DYND70802
Series is intended for use in the electromagnetic environment specified
below. The customer or the user of Medline Electrical Clipper Charging Base,
DYND70800/DYND70801/DYND70802 Series should ensure that it is used in
such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
– Guidance –
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV Contact
±15 kV Air
±8 kV Contact
±15 kV Air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast
transient/
burstIEC
61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV for
differential
mode
±1 kV for
differential
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Guidance and Manufacturer’s Declaration
– Electromagnetic Emission –
Medline Electrical Clipper Charging Base, DYND70800/DYND70801/DYND70802
Series is intended for use in the electromagnetic environment specified
below. The customer or the user of Medline Electrical Clipper Charging Base,
DYND70800/DYND70801/DYND70802 Series should ensure that it is used in
such an environment.
Emission Test
Compliance
Electromagnetic Environment
– Guidance –
RF emissions
CISPR 11
Group 1
Medline Electrical Clipper Charging
Base, DYND70800/DYND70801/
DYND70802 Series uses RF energy
only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Medline Electrical Clipper Charging
Base, DYND70800/DYND70801/
DYND70802 Series is suitable use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
RF emissions
CISPR 11
Professional
healthcare
facility
environment
Same as Group 1