Medline DYND70800 Скачать руководство пользователя страница 6

8

9

Guidance and Manufacturer’s Declaration

– Electromagnetic Immunity –

Medline Electrical Clipper Charging Base, DYND70800/DYND70801/DYND70802 
Series is intended for use in the electromagnetic environment specified 
below. The customer or the user of Medline Electrical Clipper Charging Base, 
DYND70800/DYND70801/DYND70802 Series should ensure that it is used in 
such an environment.

Immunity Test

IEC 60601 

Test Level

Compliance 

Level

Electromagnetic 

Environment

– Guidance –

Electrostatic 

discharge (ESD)

IEC 61000-4-2

±8 kV Contact

±15 kV Air

±8 kV Contact

±15 kV Air

Floors should be 

wood, concrete or 

ceramic tile. If floors 

are covered with 

synthetic material, 

the relative humidity 

should be at least 

30%.

Electrical fast 

transient/

burstIEC 

61000-4-4

±2 kV for power 

supply lines

±2 kV for power 

supply lines

Mains power quality 

should be that of a 

typical commercial or 

hospital environment.

Surge

IEC 61000-4-5

±1 kV for 

differential 

mode

±1 kV for 

differential 

mode

Mains power quality 

should be that of a 

typical commercial or 

hospital environment.

Guidance and Manufacturer’s Declaration

– Electromagnetic Emission –

Medline Electrical Clipper Charging Base, DYND70800/DYND70801/DYND70802 
Series is intended for use in the electromagnetic environment specified 
below. The customer or the user of Medline Electrical Clipper Charging Base, 
DYND70800/DYND70801/DYND70802 Series should ensure that it is used in 
such an environment.

Emission Test

Compliance

Electromagnetic Environment               

– Guidance –

RF emissions 

CISPR 11

Group 1

Medline Electrical Clipper Charging 

Base, DYND70800/DYND70801/

DYND70802 Series uses RF energy 

only for its internal function. 

Therefore, its RF emissions are very 

low and are not likely to cause any 

interference in nearby electronic 

equipment.

RF emissions 

CISPR 11

Class B

Medline Electrical Clipper Charging 

Base, DYND70800/DYND70801/

DYND70802 Series is suitable use in 

all establishments, including domestic 

establishments and those directly 

connected to the public low-voltage 

power supply network that supplies 

buildings used for domestic purposes.

Harmonic emissions 

IEC 61000-3-2

Class A

Voltage fluctuations/

flicker emissions  

IEC 61000-3-3

Complies

RF emissions 

CISPR 11

Professional 

healthcare 

facility 

environment

Same as Group 1

Содержание DYND70800

Страница 1: ...MEDLINE ELECTRICAL CLIPPER WITH CHARGING BASE INSTRUCTIONS FOR USE DYND70800 DYND70801 DYND70802 Wet Dry Charge Off O...

Страница 2: ...ade attachments are interchangeable to fit onto the Electric Surgical Clipper depending on the use of the attachment INDICATIONS FOR USE The Electric Surgical Clipper is intended for the removal of bo...

Страница 3: ...read and understand all safety information before using 1 Use this clipper only for its intended use as explained in this insert 2 The clipper must only be used with the charger provided 3 Do not sub...

Страница 4: ...en not in use after cleaning and drying place clipper body in the charging adapter for charging and safe storage Contact your Medline Representative to order replacement blades The blades are intended...

Страница 5: ...ne Electrical Clipper Charging Base DYND70800 DYND70801 DYND70802 Series should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary Medline Electrical Clipp...

Страница 6: ...ital environment Guidance and Manufacturer s Declaration Electromagnetic Emission Medline Electrical Clipper Charging Base DYND70800 DYND70801 DYND70802 Series is intended for use in the electromagnet...

Страница 7: ...maximum output power rating of the transmitter in watts W according to the transmit ter manufacture and d is the recommended separation distance in meters m Field strength from fixed RF transmitters a...

Страница 8: ...er in watts W according to the transmitter manufacturer NOTE 1 at 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situati...

Страница 9: ...ale lectrique est destin e au rasage des poils du corps des patients avant une intervention chirurgicale ou d autres op rations mineures La tondeuse peut tre utilis e de fa on s curitaire sur les poil...

Страница 10: ...ts mouvements l gers Pour obtenir de meilleurs r sultats raser dans le sens inverse de la pousse des poils Dans les endroits difficiles bien tendre la peau pendant le rasage MISES EN GARDE Toujours pr...

Страница 11: ...minutes ou une vie utile maximale de 5 ans sur le terrain Inspectez la tondeuse pour d celer tout dommage avant l utilisation et jetez la ou cessez de l utiliser si elle ne fonctionne pas Environneme...

Страница 12: ...re Si cet appareil cause des interf rences nuisibles aux autres dispositifs ce qui peut tre d termin en allumant et en teignant l appareil on encourage l utilisateur essayer de corriger ces interf ren...

Страница 13: ...riations de tension missions de scintille ment CEI 61000 3 3 Conforme missions RF CISPR 11 Environnement d tablisse ment de soins de sant professionnels Comme le groupe 1 Renseignements sur la CEM rel...

Страница 14: ...802 est con u pour tre utilis dans l environnement lectromagn tique sp cifi ci dessous Le client ou l utilisateur du chargeur de la tondeuse lectrique Medline s ries DYND70800 DYND70801 DYND70802 doit...

Страница 15: ...ieures 2 V m Directive et d claration du fabricant immunit lectromagn tique Les s ries DYND70800 DYND70801 DYND70802 sont pr vues pour tre utilis es dans l environnement lectromagn tique pr cis ci des...

Страница 16: ...l quipement de communication Puissance de sortie maximale nominale de l metteur W Distance de s paration selon la fr quence de l metteur m 150 kHz 80 MHz d 1 2 P 80 kHz 800 MHz d 1 2 P 800 kHz 2 5 GHz...

Страница 17: ...www medline com Manufactured for Fabriqu pour Medline Industries Inc Three Lakes Drive Northfield IL 60093 USA 1 800 MEDLINE Made in China Fabriqu en Chine RE21WZA...

Отзывы: