Guidance And Manufacturer’s Declaration-Electromagnetic Immunity For All Equipment And Systems
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment Guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 6 kV (contact)
± 8 kV (air)
± 6 kV (contact)
± 8 kV (air)
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Not Applicable
Not Applicable
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Not Applicable
Not Applicable
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
< 5% UT (>95% dip in UT ) for
0,5 cycle
40% UT (60% dip in UT ) for 5
cycles
70% UT (30% dip in UT ) for 25
cycles
< 5% UT (>95% dip in UT ) for
5 sec
Not Applicable
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the COMPAT
®
Standard Enteral Feeding Pump requires continued
operation during power mains interruptions, it is
recommended that the COMPAT
®
Standard Enteral Feeding
Pump be powered from an uninterruptible power supply or
a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
NOTE:
UT is the a. c. mains voltage prior to application of the test level.
ELECTROMAGNETIC IMMUNITY
The COMPAT
®
Standard Enteral Feeding Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the COMPAT
®
Standard Enteral Feeding Pump should ensure that it is used in such an environment.