Adjust paper speed:
With the speed adjusting button
on the key panel ,you can adjust
paper speed among 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s. In auto and rhythm
mode, 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s are not selectable.
Switch filter:
With the filter selection button
on the key panel ,you can switch filter
among no filter, .AC, EMG, DFT, AC+EMG, AC+DFT, EMG+DFT, AC+EMG+DFT.
AC: AC Filter
EMG: EMG Filter
DFT: Baseline Filter
Display calibration signal:
Press
on the key panel ,1mV signal will occur on the screen
once.
Print/Stop Printing:
With the printing button
on the key panel, you can begin or stop
the printing operation.
Auto Mode:
After starting printing, automatically the system prints and stores synchronous
12-lead waveforms, of which the length is according to related setup in printing settings, and also
according to this to print analysis data and conclusion, then finish printing automatically.
Manual Mode:
After starting printing, the user can print different lead instant waveforms by
switching the lead displayed, namely the ECG printed in manual mode is not synchronous and the
data will not be saved. You need to stop printing by pressing
again.
During printing process, the printing state and display content mainly include:
Display content
Explanation
Process…
It is printing.
Waiting…
It is finishing printing.
No Paper.
Paper lack, the user should restart printing after loading paper.
Print Timeout.
Communication failure between this system and printing sub-system.
ECG Timeout
Communication failure between this system and sampling sub-system.
Low Power
Low power, it can not start printing
Note:
You can not print until ECG waveforms are displayed on the screen.
Here in this interface, press
to quick enter quick setup interface showed below.