67
When the probe connected to the system, do not clean or disinfect the
probe.
Do not use gasoline or other organic compound solution to clean the
probe.
If the probe sticks to body fluids, please disinfect and then clean.
After each ultrasonic examination, completely erase the probe surface of
the ultrasonic coupling agent. Otherwise the ultrasound coupling agent
will coagulate in the probe acoustic lens and affect the ultrasonic image
quality.
Clean the probe regularly to ensure that the probe is working properly.
Cleaning step
Make sure that the probe cable is not cracked or ruptured;
Wipe the dust from the cable with a soft cloth.
Gently wipe the probe and cable with a soft cloth soaked in an aqueous
solution and a mild detergent;
Use a soft cloth to wipe the coupling agent attached to the probe;
Wipe the probe and cable with a soft cloth soaked in warm soapy water;
Wipe the probe and cable with a soft cloth dampened in fresh water until
all soap liquid is erased;
Dry the probe and cable with a dry soft cloth.
11.4.3
Sterilization of the probe
Sterilization of the probe
The ultrasonic probe can be sterilized with a liquid chemical disinfectant.
The degree of disinfection is directly related to the time of exposure to the
disinfectant. Increase the contact time has a better disinfection effect.