Error Messages
MESSAGE
WHAT IT MEANS
WHAT TO DO
E-b
Appears when the battery
is too low.
Replace the battery immediately.
E-U
Appears when a used
test strip is inserted.
Repeat with a new test strip.
E-t
Appears when ambient
temperature is above or
below system operation
range.
System operation range is 8°C to
45°C (46.4°F to 113°F). Repeat the
test after the meter and test strip are
in the above temperature range.
E-0
E-A
E-E
Problem with the meter.
Repeat the test with a new test strip.
If the meter still does not work,
please contact the customer service
for assistance.
E-F
Appears when test strip is
removed while counting
down, or insufficient
blood volume.
Review the instructions and repeat
test with a new strip.
If the problem persists, please
contact the local customer service
for help.