22
For the lancing device:
Take out one disinfecting wipe from the container.
Clean exposed surfaces of the lancing device thoroughly and
remove any visible dirt, blood, or any other body fluid with the
wipe.
Take out second disinfecting wipe from the container. Keep the
surface of lancing device wet with disinfection solution for a
minimum of 1 minute.
Allow the surface to dry completely.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact customer service
or place of purchase for assistance.
This device has been validated to withstand 260 cycles (10,950
cleans), of cleaning and disinfection, using the recommended
disinfecting wipe, to simulate one cleaning and disinfection cycle per
week over 5 years of use.
The lancing device has been validated to withstand 110 cycles of
cleaning and disinfection, using the recommended disinfecting wipe,
to simulate one cleaning and disinfection cycle per week over 2
years of use.
The meter should be replaced after the validated number of cleaning
and disinfection cycles or the warranty period, whichever comes
first. Stop using the meter if you see any signs of deterioration. For
example:
meter cannot be turned on
LCD display cracks or becomes cloudy
buttons no longer function
Содержание DGEMP01M
Страница 1: ...Blood Glucose Monitoring System DGEMP01M Owner s Manual Owner s Manual ...
Страница 2: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 28: ...17 WARNING While the meter is connecting to the PC it will be unable to perform a blood glucose test ...
Страница 42: ......
Страница 43: ......