The battery change symbol appears in the display
• The batteries are too low or empty. Replace both batteries with new 1.5 V AAA LRO3 batteries.
Erratic measurements being displayed
• Cuff incorrectly fitted or not tight enough. Fit cuff correctly and tighten.
• You have changed position or the cuff has moved during the measurement. Keep still and do not
move your arm or the cuff.
• You have an irregular heartbeat. Even if your heartbeat is only slightly irregular, repeat the
measurement. The unit is not suitable for people with an extremely irregular heartbeat.
• You have spoken during the measurement, you are nervous or you are excited. Do not speak and
breathe deeply two or three times in order to relax.
• You are not in the corrected seated position. Please assume the correct position when taking the
measurement.
• A problem occurred during inflation or during the measurement. Repeat the measurement.
“EE” appears in the display
• The unit has not been inflated enough. Repeat the measurement
• A problem occurred during inflation or during the measurement. Repeat the measurement.
No display or display locked
• No batteries have been inserted, batteries wrongly inserted or batteries are completely empty.
Insert batteries, check battery position or replace with new batteries.
• A malfunction or fault has occurred. Briefly remove the batteries and then replace them.
Please contact the service centre if the error message appears again when the measure-
ment is repeated.
31
GB
6 Miscellaneous
51066/51067_HGN_Mini-Format_West.qxd:Bltudruckmessgerät 01.02.2012 16:28 Uhr Seite 31
Содержание 51066
Страница 168: ...162 GR 1 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 29 Uhr Seite 162...
Страница 169: ...IIa B LOT 1 1 163 GR 1 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 29 Uhr Seite 163...
Страница 170: ...START 10 1 2 164 GR 1 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 29 Uhr Seite 164...
Страница 174: ...5 2 7 168 GR 2 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 30 Uhr Seite 168...
Страница 175: ...2 1 5 V AAA LR03 3 1 169 GR 3 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 30 Uhr Seite 169...
Страница 180: ...6 1 1 5 V A LR03 174 GR 6 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 30 Uhr Seite 174...
Страница 181: ...6 2 6 3 175 GR 6 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 30 Uhr Seite 175...
Страница 184: ...1 MEDISANA 2 3 4 5 178 GR 7 51066 51067_HGN_Mini Format_West qxd Bltudruckmessger t 01 02 2012 16 30 Uhr Seite 178...