Attention
The medinSINDI
®
data interface must be connected using the cable specified above and a computer system that
meets the medically relevant requirements in terms of electrical safety and EMC, otherwise the functionality
and EMC of the medinSINDI
®
cannot be guaranteed.
4.4.5. Illumination of the Flowmeter
For medinSINDI
®
units built in 2009 or later, the flowmeter display is illuminated, allowing the amount of flow
set to also be read in darkened rooms.
For older medinSINDI
®
s, this illumination can be retrofitted if needed.
5.
Alarm
medinSINDI
®
differentiates between two different alarm centres: the gas mixer and the CPU monitor.
5.1.
Gas Mixer
Pneumatic-acoustic alarm: is triggered independently of the electronics if one of the two supply gases is
missing (air/O2) or the pressure difference between air and O
2
is greater than 1 bar.
5.2.
Monitoring Alarms in the medinSINDI
®
CPU Monitor:
Fix Levels
-
Low CPAP Alarm (CPAP ≤ 1.5 cm/H
2
O): Disconnection alarm – check all tube connections.
-
High CPAP Alarm (CPAP ≥ 10 cm/H
2
O): Overpressure alarm – check flow settings and tubes
Setting Levels
-
Low CPAP: Monitoring of the limit values set (up to software version 1.015, activation necessary
"CONFIRM CPAP")
-
High CPAP: Monitoring of the limit values set (up to software version 1.015, activation necessary
"CONFIRM CPAP")
Low or High Oxygen Concentration
Monitoring of the oxygen concentration set. Activation with "CONFIRM FiO2" necessary
Battery
In the event of a shortfall of the minimum charge, there will be a 1-second beep each minute