IXL4L_IXX3L07/08/11
MEDICATLANTIC
2015_10_15
7/32
If the side rail is poorly positioned, safety of the patient may be endangered or a
malfunctioning may occur. The side rails must not be used when the patient is a child (under
12) or if s/he is too small
(
≤
146 cm
).
Check that the side rail is locked by trying to fold it without using the release pin.
There must be at least 220 mm between the top of the side rail and uncompressed mattress
surface.
3.3.2. Metal side rails
To positionning the side rail.
1
2
1 Insert amount of the side rail in the grip jaw on the head side of the bed.
2 Tighten the Rondo screws of the grip jaws on the bed base.
To remove the side rail, in reverse operation 2 and operation 1 by pulling the release pin on the grip jaw of
the head side.
To lower the side rail.
To raise the side rail.
Head side
Feet side
Feet side
Head side