EasyOne Pro™ / EasyOne Pro™ LAB
89
Hygiene, Cleaning, Maintenance, Disposal
7
Hygiene, Cleaning, Maintenance, Disposal
Hygiene
Caution
Risk of Infection—
Users with open wounds or contagious diseases on their
hands must wear gloves.
Wash your hands after each patient to protect yourself
and to prevent cross contamination.
The
spirette™
and the
barriette™
must be properly
disposed of immediately after use.
Cleaning
Warning
Shock Hazard—Disconnect the device from the power line
before cleaning.
Wipe the device surface down with a cloth moistened with alcohol (70%
max.), avoiding the display, the rear panel and the temperature and
humidity sensor. Do not let liquid enter the device.
To clean the rear panel and the temperature and humidity sensor, simply
wipe them off with a dry cloth.
The display is also cleaned with a soft, dry cloth. In case of major
contamination, you can use a cloth moistened with alcohol (50% max.).
Wipe the sensor, the valve unit, the gas supply tube and the power cord
down with a cloth moistened with alcohol (70% max.). Do not let liquid
enter the parts. Replace the gas supply tube each year.
The connecting tube to the gas cylinder and the touch stylus can be
cleaned with a cloth moistened with alcohol (70% max.).
The one-way valve can be rubbed down with a cloth moistened with
alcohol (50% max.). Clean only the side facing the
barriette™
and the
housing. The one-way valve must be replaced each year.