EasyOne Pro™ / EasyOne Pro™ LAB
25
Start-up and Initial Preparation
Connect the temperature and humidity sensor.
Note
Ensure that the sensor is exposed to the normal room
temperature and that it will not measure incorrect values
because of exposure to a draft from the blower, to lamps or
direct sunlight.
2.4 Gas Supply
DLCO Gas
Note
It is very important to comply with the information given in
chapter
The gas cylinder fittings (pressure relief valve) must meet the
following minimum requirements:
input pressure on primary side: 150 to 300 bar (depending on
max. cylinder pressure)
output pressure on secondary side: 0 to 6 bar
flow rate: 6 L/s minimum
You can purchase all necessary gases and cylinders as well
as the fittings from your local ndd dealer.
Equipment not intended for mobile use can also be used
with large gas cylinders.
Use only the supplied original tubing to connect the gas
cylinder to the device.
Use only the original gas supply tube to connect the valve
unit to the device.
Fig. 2-9
Temperature and humidity sensor