STEP 3.
Insert the electrodes in the cap accordance with the research scheme.
Point EEG electrodes insert into fixing ring until it stops.
Cup EEG electrodes insert into electrode hole.
Pay attention that electrode cables are not confused and not bended.
To avoid confusion of wires, use coupler for wires. Insert a coupler tip through the hole at
the other end, strongly tighten and then fix a tip with velcro.
STEP 4.
Connect the electrodes to the skin through EEG gel.
Fill the syringe with gel.
Gel should not be too fluid nor too viscous.
Set supplied the nozzle-
dispenser. The nozzle-dispenser should sit tightly on the syringe and should not hang or
subside. Use only supplied the nozzle-dispenser with blunt needle. Spiky needle can
damage patient skin.
Slide the hair directly under the EEG electrodes with use the nozzle-dispenser needle. It will
help to reduce the impedance and increase the quality of researches.
STEP 5.
Fast connection.
In cases where it is necessary to promptly fast disconnect or connect the electrodes
mounted on the patient, for example, when registering somnogrammy, can be used passive
adapter. It can also be used when it is necessary to increase the distance from the patient
to the amplifier.
Minimizing impedances
Recommended to use electroencephalograph in mode of impedance measurement.
Insert a syringe needle of the nozzle-dispenser into a hole of electrode and squeeze out a
few gel. Then take out a needle from hole and remove remaining gel. If it has not given
results repeat procedure. If after that the impedance remains too high,
check the electrode.