INSTRUCTIONS FOR USE

 

STEP 1. 

Measure the circumference of the patient's head.  

Head size measure around the fullest part of the head just above the ears using a flexible tape 
measure.  Taking  the  measurement  at  the  forehead.  Circumference  of  the  head  should  be 
measured at a point approximately one inch above the eyebrows or at whatever point gives the 
largest possible measurement. After the measuring choose the proper cap size. 

STEP 2. 

Put the cap on the patient.  

Beginning from the front of head, where the subject may hold the edge of cap, pulling the cap 
over of the head. Next, adjust the electrode position 

Cz 

halfway between Nasion and Inion. The 

frontopolar and occipital electrode positions should now be placed correctly. Otherwise choose 
another cap size. When measured correctly, the frontopolar electrodes (e.g. 

Fp1/Fp2

) should lie 

directly above the eye brows. 
Make  sure,  that 

Cz

  is  right-left-centered  as  well,  and  that  lateral  electrode  positions  are 

symmetrical. 
Affix the cap with the chin belt. Pay attention to the correct location of its. The wide part of a belt 
should fit a chin and shouldn’t slump on the neck. The belt tension should provide reliable fixing 
of cap. 

Don’t pull in tight chin belt. The subject should feel comfortable even during long-term research.

 

 

Attach the cap on the head is possible with use the chest belt. This type of attachment is 
effective in cases where during the test pressure of the chin strap hinders the patient's actions, 
for example, if the patient should talk. 
The chest belt is attached directly to the patient's chest. Chest belt is arranged in V-shape or X-
shape, which contributes to a secure fit of the cap. Tension is regulated by the length of the 
straps.

 

Содержание MCScap Clinic

Страница 1: ...1993 2018 Medical Computer Systems Ltd MCScap EEG electrode caps USER MANUAL...

Страница 2: ...fic researches GMDN code 63131 Electroencephalographic electrode cap reusable Product code by 93 42 EEC MD 0100 General non active non implantable medical devices Classification class I as per Appendi...

Страница 3: ...s to any part of the product including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result If such use is necessary this equipment and the othe...

Страница 4: ...d service Please look thoroughly through this manual before its use The manual contains detailed information and guidelines required for proper and safe operation of this product Failure to follow the...

Страница 5: ...belt Reusable EEG electrodes have a channel for gel and the recording surface is silver silver chloride Ag AgCl Sintered technology of sensors guaranties minimal polarization and long term stability E...

Страница 6: ...and Inion The frontopolar and occipital electrode positions should now be placed correctly Otherwise choose another cap size When measured correctly the frontopolar electrodes e g Fp1 Fp2 should lie d...

Страница 7: ...ozzle dispenser should sit tightly on the syringe and should not hang or subside Use only supplied the nozzle dispenser with blunt needle Spiky needle can damage patient skin Slide the hair directly u...

Страница 8: ...pull the electrode Under most circumstances cleaning of caps and electrodes is carried out in a detergent Be attentive dish detergents often leave a film A children s shampoo will be a good choice Im...

Страница 9: ...nectors do not comply with the cap Replace the electrode or the cap Electrode connectors can not be connected to the electroencephalograph Electrode connectors do not comply with the electroencephalog...

Страница 10: ...scientific research high resolution EEG MR compatible EEG available by order routine EEG scientific research Electrode type MCScap MCScap T MCScap NTC MCScap NT available by order MCScap E MCScap T M...

Страница 11: ...ng on the type of electrode The impedance of the electrode 5 k 5 k 5 k 5 k Spare electrode no no 1 1 Channels in Standard caps Fp1 Fp2 F3 F4 C3 C4 P3 P4 O1 O2 F7 F8 3 T4 T5 T6 Cz Fz Pz GND Fp1 Fp2 F3...

Страница 12: ...G starter kit gel syringe needle brush bag User manual 2 textile caps MCScap 10 20 with fixing rings 26 EEG electrodes MCScap E ear electrode fixators 2 sets of labels for marking electrodes EEG start...

Страница 13: ...ghted in bold font 2 Cycle refers a use disassembly washing and disinfection of the helmet cleaning and disinfection of the electrodes Compliance Restriction of Hazardous Substances The object is in c...

Страница 14: ...460 Russian Federation www mks ru Page 14 from 16 10 20 system Electrode positions and labels yellow circles indicate positions of the 10 20 system T3 T4 T5 and T6 corresponds to T7 T8 P7 and P8 of th...

Страница 15: ...n Federation www mks ru Page 15 from 16 10 10 system Electrode positions and labels yellow circles indicate positions of the original 10 20 system white circles indicate additional positions introduce...

Страница 16: ...MCScap size table Correct size is important for quality EEG acquisition and patient comfort Check the head circumference of patient before EEG examination...

Отзывы: