WARNING
Do not use the EEG cap with removable passive Ag/AgCl electrodes and accessories (hereafter - product) for other
purposes than it is intended.
Use IEC 60601-1 standard to combine the product with other devices such as encephalograph. The operator bears
the responsibility for use of any devices together with a product. Use of a product with other devices should not lead
to failure and-or increase of a risk level of the operator and the patient. Using electrodes of the same type.
Do not operate the product within powerful radio interference producing sources such as arc welders, radio thermal
treatment equipment, x-ray machines or any other equipment that produces electrical sparks. Portable RF
communications equipment (including peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the product, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
Do not use in magnetic resonance imaging.
Always observe the safety requirements of this manual for operation and maintenance. If any problems don’t open the
product, do not unauthorized repairs, any doubts please contact your local supplier.
Take care of the product’s components. Avoid the cables’ breaks, kinks, tension, excessive stretching of the EEG cap
and other mechanical efforts. Take care of arranging cables/wires to avoid the risk of entanglement or strangulation.
Electrode sensors are fragile and require careful and caring treatment. Always put electrodes on a soft surface, such
as a napkin or a towel.
Reapply of a product present a potential risk of cross-infection especially when are used on abraded skin, unless they
are restricted to a single patient or disinfection between patients.
Use the special EEG electroconductive gel/paste for the EEG acquisition.
Don’t break the conditions of transportation, storage and operating of the product. Do not expose the product to direct
sunlight.