![Mediana HeartOn T15 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/mediana/hearton-t15/hearton-t15_operators-manual_1760605011.webp)
9
Custom Scenario
The default custom scenario is simulated if the custom scenario setting is not changed by
user.
Changing Custom Scenario
1. Open the external DIP switch box at bottom panel.
2. Place the 5 and 6 pin of the
External DIP Switch
to down position to activate custom
scenario function.
3. Connect the USB mini B cable.
4. Run the custom scenario software (option) on the PC.
5. Set up the serial port (COM port) on the PC.
6. Select the desired scenario and then perform the download to the custom scenario 1, 2,
or 3.
7. After completion of downloading, turn off the AED by closing the AED cover and
disconnect the USB mini B cable.
8. Set the
External DIP Switch
to normal operation.
9. Turn on the AED by opening the AED cover again and verify the setting of custom
scenario by pressing the
Custom Scenario Button
on the remote control.
Table 5. Setting of External DIP Switch
Setting of External DIP Switch
Description
Setting available
Normal operation
Standard use
Changing custom
scenario
Standard use
Changing audible
list
Not use without approval by
service personnel