Using the AED Trainer
Operating the AED Trainer
1. Push the
Slide Button
to open the AED cover.
2. If the AED is turned on by pushing the
Slide Button
, the first scenario is started
automatically.
3. To change the scenario during other scenario simulation, press the desired scenario
number on the remote control then a selected scenario will be started automatically.
4. The volume can be controlled by 10 steps. The last setting will be stored and maintained
even if the AED has been turned off.
Function of remote control
The shockable rhythm is simulated to deliver the shock immediately. After
deliver the shock, the next scenario is simulated.
The non-shockable rhythm is simulated. The CPR scenario is simulated
without shock delivery.
The poor pad condition is recovered to the good pad condition. After then,
the AED Trainer resumes on the analyzing rhythm scenario.
The motion artifact is simulated.
When the button is pushed once, the poor pad condition is simulated and
push again, the changing pad condition is simulated. When the pad
disconnects and connects again, the changing pad condition is recovered
to normal condition.
Not used in HeartOn T15 and HeartOn T15-G4.
Simulate the Low battery condition and indicate Low battery condition on
the status indicator. When the button is pushed again, the AED Trainer
displays real battery level on the status indicator.
Pressing one time, the AED is operated normally, current temperature is
indicated low, and temperature icon is flashing. Pressing once again, error
condition is simulated, the X mark is displayed on the status indicator and
temperature icon is lit. Finally, pressing once again, the AED will not be
operated, unless the AED cover is closed and then reopened.
Resume the audible activity from the point at which it was paused.
Pause all audible activity in the AED Trainer. If
button is not pushed,
the AED Trainer does not play any audible activity during pushing any
button.
The starting point of the standard scenario can be selected by user.
MEDIANA OPTION button
is toggles between starting point from the
checking patient’s consciousness and starting point from the beginning
part. (HeartOn T15 only. The
MEDIANA OPTION button
press is ignored
on HeartOn T15-G4.)
Note: When the pads are not connected during scenario simulation, the audio prompt
sounds repeatedly to ask the pads is inserted.