29
IX. Programs in the Amadé
For different listening situations, up to three programs
can be activated. All Amadé parameters can be
modified and individually adjusted in every listening
program. The programs can be seen sorted 1 to 3
from top to bottom either on the right or the left
margin of the Connexx programming screen (figure 32).
Presets available in the
Amadé
The following presets are available: Universal, Noisy
Environment, TV/Music, Telephone acoustically, Outdoor/
Sports, and Privacy.
Universal
The “Universal” program, which is stored under slot 1,
is automatically active after switching on the Amadé.
The microphone system is set to omnidirectional.
Speech and Noise Management as well as Sound
Smoothing are set to medium, and the Wind-Noise
Reduction is set to maximum.
(Please find more information about these features on
pages 26 and 27 of this guide).
In slot 1 the program Universal is always active.
Nevertheless you can change any setting within the
program.
Noisy Environment
The program “Noisy Environment” is predefined in slot
2 and designed for noisy situations such as traffic,
conversation in a large group, restaurants, or similar
noisy situations.
To provide speech intelligibility and comfort within
these situations the Amadé parameters are modified
as follows: the microphone system is set to directional.
Additionally Speech and Noise Management is set to
maximum attenuation, Sound Smoothing is set to
medium, and the Wind-Noise Reduction is set to
maximum. Furthermore the channel gains are reduced
in the lower frequency range as well as the gain for
loud sounds for the whole frequency range.
TV/Music
The “TV/Music” program, which is the predefined
program for slot 3, is primarily intended for listening
to music but also for TV applications. It aims to provide
a natural, wide band sound for the Amadé user. To
achieve this goal, Wind-Noise reduction is turned off
and for this particular acoustical environment, the
Sound Smoothing feature is also turned off. The overall
gain is also increased, especially in the low and high
frequency ranges, to provide more bandwidth and
sound quality. The time constants of the compression
are set to high values (dual) and the microphone
system is set to omnidirectional mode. Speech and
Noise Management is set to medium attenuation.