SpiroConnect Operators Manual
MedChip Solutions Ltd.
Page 23 of 31
001.105.GBR Issue 1.12
10
Looking after your SpiroConnect
Please observe the following precautions:
•
Avoid exposing the SpiroConnect to direct sunlight.
•
Avoid operating the spirometer in dusty conditions or near to heating appliances or
radiators.
•
Do not keep the spirometer in a damp place or expose it to extremes of temperature.
11
Cleaning
11.1
Unit Case Cleaning COVID-19 Update
After each use the unit should be wiped down with an alcohol infused wipe or other
medically approved anti-viral wipes such as Clinell wipes.
See also section
“1.1 MedChip Solutions Covid-19 Response”.
Take care that no liquid is allowed to enter the unit.
11.2
Turbine/Transducer Cleaning
The transducer requires no routine maintenance or servicing. To clean or disinfect the
transducer it may be removed by means of the following procedure:
1
Remove the transducer by rotating the mouthpiece holder assembly anti clockwise by
90
o
and gently pulling from the main body.
2
The transducer may now be immersed in warm soapy water for routine cleaning or
immersed in Rely+On
TM
Perasafe
TM
solution for a period not exceeding 10 minutes. After
cleaning/disinfection, the transducer should be rinsed in distilled water and dried.
CAUTION: Solutions containing alcohol or chlorine, e.g. bleach, must not be used.
3
Re-assemble the transducer into the main body by reversing the steps shown for
disassembly.
CAUTION: When removing or replacing the turbine, the mouthpiece holder must be
facing away from the SpiroConnect product name as shown above. Do not use
excessive force.
11.3
Soft Carry Case Cleaning
Local infection control guidance should be followed. Alcohol infused wipe or other medically
approved anti-viral wipes may be used.