Maintenance
Repairs
26
GA 5752 4981 GB 15
7
7.4
Repairs
The following issues may require repairs to be made by the manufacturer or an authorised
service partner:
• Liquid has penetrated the device.
• The performance has significantly decreased.
• Inexplicable notifications appear.
• Abnormal noises occur.
• Functional faults cannot be rectified according to the measures in chapter Malfunctions and
If defects are detected, the product may not be used any longer.
Make a note of the deficiencies and the REF number on the data plate and inform the responsible
ATMOS Service.
Observe the information in chapter Sending in the device [
7.5
Service hotline
+49 7653 689-0
7.6
Type plate position
1
Fig. 7: Type plate
Position of the type plate (1).
7.7
Sending in the device
Remove and properly dispose of consumables.
Clean and disinfect the product and accessories according to the operating instructions.
Place used accessories with the product.
Fill in the form QD 434 ‘Delivery complaint / return shipment’ and the respective
decontamination certificate
.
This form is enclosed with each delivery and can be found at www.atmosmed.com.
The device must be well padded and packed in suitable packaging.
Place the form QD 434 ‘Delivery complaint / return shipment’ and the respective
decontamination certificate
in an envelope.
Affix the envelope to the outside of the package.
Send the product to ATMOS or to your dealer.