AJUSTING THE CONTROLS USING
MANUAL ADJUSTMENTS
Choose the desired mode by pressing the mode Mode Button
There are 3 modes available Constant, Synchronous and Alternate.
Constant
Constant or normal is stimulation based on setting value.
Only pulse width, pulse rate and timer are adjustable in this
mode.
Synchronous
Stimulation of both channels occurs synchronously. The
"ON" time including "Ramp Up" and "Ramp Down" time.
Therefore, the setting of ON Time should be no less than
two times of the Ramp Time in this mode.
Alternate
The stimulation of CH2 will occur after the 1st contraction
of CH1is completed. In this mode, the setting of the ON
Time should be no less than two times of the Ramp time.
The Off Time should be Equal or more than the On Time.
When you have chosen the desired mode now press the SET button, toggle
through each parameter, you can enter the required setting. You can
change the value by pressing the Increment UP and Down controls when
the value is flashing.
In Constant Mode, you can change the Pulse Width, Pulse Rate and timer.
In Synchronous Mode, you can change the Pulse Width, Pulse Rate, Ramp
Time On, Off and timer.
In Alternate Mode, you can change the Pulse Width, Pulse Rate, Ramp Time
On, Off and timer.
CH1
Constant
CH2
CH1
Synchronous
CH2
CH1
Alternate
CH2