AJUSTING THE CONTROLS USING
MANUAL ADJUSTMENTS

Choose the desired mode by pressing the mode Mode Button

There are 3 modes available Constant, Synchronous and Alternate.

Constant

Constant or normal is stimulation based on setting value.
Only pulse width, pulse rate and timer are adjustable in this
mode. 

Synchronous

Stimulation of both channels occurs synchronously. The
"ON" time including "Ramp Up" and "Ramp Down" time.
Therefore, the setting of ON Time should be no less than 
two times of the Ramp Time in this mode.

Alternate 

The stimulation of CH2 will occur after the 1st contraction
of CH1is completed. In this mode, the setting of the ON
Time should be no less than two times of the Ramp time.
The Off Time should be Equal or more than the On Time.

When you have chosen the desired mode now press the SET button, toggle
through each parameter, you can enter the required setting. You can
change the value by pressing the Increment UP and Down controls when
the value is flashing.

In Constant Mode, you can change the Pulse Width, Pulse Rate and timer.

In Synchronous Mode, you can change the Pulse Width, Pulse Rate, Ramp
Time On, Off and timer.

In Alternate Mode, you can change the Pulse Width, Pulse Rate, Ramp Time
On, Off and timer.

CH1
Constant

CH2

CH1

Synchronous

CH2

CH1

Alternate 

CH2

Содержание Comfy EMS Plus

Страница 1: ...se read this User Manual before using your Device Tel 0161 429 7330 www tensmachineuk com email sales thetenscompany co uk The Quick and Easy User Guide for the Maxi Stim EMS Plus Neuromuscular Stimul...

Страница 2: ...ompany registration number 08758741 Vat registration number 308286105 Thank you for purchasing a Med Fit Maxi EMS Plus Stimulator This device is manufactured to the highest of medical standards using...

Страница 3: ...STEP GUIDE 4 MUSCLE STIMULATOR PROGRAMMES 7 MANUAL ADJUSTMENTS 8 INTRODUCTION 11 HOW EMS WORKS 11 CAUTIONS 12 CONTRAINDICATIONS 12 ADVERSE REACTIONS 12 WARNINGS 13 ELECTRDOE INSTRUCTIONS 15 EMS PROGRA...

Страница 4: ...mpartment at the rear of the device This easy to use instruction manual explains how to set up the Med Fit Maxi Stim EMS Plus in programme mode Battery compartment Belt clip Fig A STEP BY STEP GUIDE 4...

Страница 5: ...nge the programme simply press the up down keys to the desired programme To access the controls open the front cover opens from the left hand side see Fig B for control functions Easy to use 2 channel...

Страница 6: ...ts as shown in Fig C Depending on the condition and treatment area you may use any number of electrodes up to a maximum of 4 electrodes and 2 channels Increase the intensity to the desired level pleas...

Страница 7: ...engthening Muscle Re Education Programmes Choose from one fo the programmes 5 6 7 and 8 Suitable Conditions These programmes can be used for building stamina and strength Muscle re eduction Please fol...

Страница 8: ...ernate The stimulation of CH2 will occur after the 1st contraction of CH1is completed In this mode the setting of the ON Time should be no less than two times of the Ramp time The Off Time should be E...

Страница 9: ...uld be selected for muscle strengthening and muscle bulk 50 120Hz pulse per second This range of frequency is used for strengthening re education and spasticity Using higher frequencies 50Hz plus requ...

Страница 10: ...ually 2 4 seconds ramp up time and 1 2 seconds ramp down time For a stronger stimulation increase ramp up time to maximum 8 seconds Stimulation modes Continuous mode device runs continually Synchronou...

Страница 11: ...ings This is being widely used in hospitals and sports clinics for the treatment of muscular injuries and for the re education of paralyzed muscles to prevent atrophy in affected muscles and improving...

Страница 12: ...lectrical conductive medium The irritation can usually be reduced by using an alternate conductive medium or alternate electrode placement 6 Electrode placement and stimulation settings should be base...

Страница 13: ...y to the carotid sinus reflex 3 Stimulation should not be applied over the neck or mouth Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to clo...

Страница 14: ...ent acid may leak from used batteries and damage the device Put the device and accessories in carrying box and keep it in cool dry place 6 The packed TENS EMS device should be stored and transported u...

Страница 15: ...ree environment until the gel regains its tack Removal and storage 1 Lift a corner of the electrode and slowly peel the electrode off the skin touching the adhesive gel as little as possible 2 Place t...

Страница 16: ...ec sec min 1 Muscle Training Re Education SYNCHRONOUS 30 300 0 1 1 30 2 Muscle Training Re Education SYNCHRONOUS 30 300 1 5 2 30 3 Muscle Training Re Education SYNCHRONOUS 30 300 1 10 3 30 4 Muscle Tr...

Страница 17: ...t areas of the body We are definitely not trying to give an anatomy class here but is is necessary to understand the whole muscle system to get a more precise grasp of the human muscle system The musc...

Страница 18: ...18 www tensmachineuk com Telephone 0161 429 7330 ELECTRODE PLACEMENT Forearm Biceps...

Страница 19: ...for muscle training and for gaining volume using the devices muscle strengthening programmes But it is equally very useful for diminishing the effects of lactic acid substances manufactured by the mus...

Страница 20: ...20 Telephone 0161 429 7330 www tensmachineuk com ELECTRODE PLACEMENT Internal Thigh Thigh...

Страница 21: ...21 www tensmachineuk com Telephone 0161 429 7330 The Gluteals This position is ideal for shaping the gluteal muscle ELECTRODE PLACEMENT The Trapezius muscles and Dorsals...

Страница 22: ...22 Telephone 0161 429 7330 www tensmachineuk com ELECTRODE PLACEMENT The Legs and Calves The medial and lateral gastrocnemius...

Страница 23: ...Telephone 0161 429 7330 The Abdominals It is the abdominal muscle which when electro stimulated will make a six pack appear ELECTRODE PLACEMENT The Pectorals BE SURE to follow the positioning DO NOT P...

Страница 24: ...s Mode Stimulation of both channels occurs synchronously The ON time including Ramp Up and Ramp Down time Therefore the setting of ON Time should be no less than two times of the Ramp time in this mod...

Страница 25: ...ality should be IEC 61000 4 5 line s mode that of a typical commercial or hospital environment Voltage dips 5 UT 5 UT Mains power quality should be short 95 dip in UT 95 dip in UT that of a typical co...

Страница 26: ...xed RF transmitters as determined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with...

Страница 27: ...ording to frequency of transmitter output power of m transmitter W 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 68...

Страница 28: ...Tel 0161 429 7330 Fax 0161 427 0215 www tensmachineuk com...

Отзывы: