24
Telephone: 0161 429 7330 www.tensmachineuk.com
TECHNICAL SPECIFICATIONS
MECHANISM TECHNICAL DESCRIPTION
01
Channel
Dual, isolated between channels
02
Pulse Amplitude
Adjustable, 0-100 mA peak into 500 ohm load each channel.
03
Wave Form
Asymmetrical Bi-Phasic Square Pulse
04
Voltage
0 to 50V (Load: 500 ohm)
05
Power source
One 9 Volt Battery
06
Size
10.1cm(L) x 6.1cm(W) x 2.45cm(H)
07
Weight
150 grams with battery.
08
Pulse Rate
Adjustable, from 2 to 150 Hz, 1 Hz/step
09
Pulse Width
Adjustable, from 50 to 300 microseconds, 10 us/step
10
On Time
Adjustable, 2~90 seconds , 1 Sec./ step
11
Off Time
Adjustable, 0~90 seconds , 1 Sec./ step
12
Ramp Time
Adjustable, 1~8 seconds, 1 Sec./ step, The “On” time will increase and
decrease in the setting value.
13 Output Mode
Constant, Synchronous, Alternate
14 Constant
Constant stimulation based on setting vak=lue. Only pulse wudth, pulse
rate and timer are adjustable in this mode.
15
Synchronous Mode
Stimulation of both channels occurs synchronously. The “ON” time
including “Ramp Up” and “Ramp Down” time. Therefore, the setting of
ON Time should be no less than two times of the “Ramp” time in this
mode. ON TIME ≥ Ramp up + Ramp down.
16
Alternate Mode
The stimulation of the CH2 will occur after the 1st contraction of CH1 is
completed. In this mode, the setting of ON Time should be no less than
two times of the “Ramp” time. The OFF Time should be equal or more
than the ON Time. ON TIME ≥Ramp up + Ramp down OFF TIME ≥ON
TIME
17
Timer
Adjustable, from 1 to 60 minutes or Continuous. Adjustable in 1 minute
each step from 1 to 15 minutes, and 5 minutes each step from 15 to 60
minutes. Treatment time countdown automatically.
18
Patient Compliance Meter
This unit can store 60 sets of operation records. Total recorded time is 999
hours.
19
Low Battery Indicator
A low battery indicator will show up when the battery is low.
20
Operating Condition
Temperature:10°~40°C Relative Humidity: 30%~75%
Atmosphere Pressure : 700Hpa~1013Hpa
25
Remark
There may be up to a +/-5% tolerance of all parameters and +/-20%
tolerance of output amplitude & voltage.