CHAPTER 1
| INTRODUCTION
The ENV-597 Water Maze Pool Heater is intended to be used to maintain a constant
temperature in either the mouse- or rat-sized pools. The heater will arrive installed in the water
maze pool configured for ˚C display. The user will set their desired pool temperature. This
manual should be carefully read prior to operating the heater.
Figure 1.1 – Water Maze Pool Heater
Warni ngs
Note:
Failure to comply with the following actions will result in heater failure and possible
equipment damage, injury or fire.
1.
Place the heater controller in a dry location
. Moisture in the controller will cause failure.
2.
Do not operate the heater when it is out of the water or in low water
. An exposed heater
is a potential fire hazard.
3.
Allow the heater to cool for 20 minutes before removing it from the pool
.
4.
Protect the heater from power surges by having the properly sized circuit breaker or inline
fuse. See label on heater controller.
5.
Be sure the thermostat sensor is immersed in water and positioned away from the heater