26
Peristaltic pump tubing
It is advisable do not exceed 8 sterilization
cycles
Altitude
Less than or equal to 2000 meters
Weight and size
3,2Kg
300 x 250 x 95 mm (L x l x H)
Table 2 – Technical Specifications
a)
W = width ; L = length ; H = height
8.1
Electromagnetic Compatibility IEC/EN 60601-1-2
WARNING:
Contraindications.
Interference with other equipment.
Though compliant with the standard IEC
60601-1-2, PIEZOSURGERY® GP may
nonetheless interfere with other devices
nearby. PIEZOSURGERY® GP must not be
used near to or stacked on other devices.
If adjacent or stacked use is necessary,
PIEZOSURGERY® GP and the other devices
must be observed and checked to verify
normal operation in the configuration in
which they will be used.
WARNING:
Portable and mobile radio
communication appliances may affect proper
device operation.
WARNING:
Contraindications.
Interference from other equipment.
An
electrical scalpel or other electro-surgical
units near the PIEZOSURGERY® GP device
may interfere with its correct operation.
WARNING:
The device requires specific
EMC precautions and must be installed
and commissioned according to the EMC
information provided in this chapter.
WARNING:
Only use original
Piezosurgery/Mectron accessories and spare
parts. The use of cables and accessories not
supplied by Piezosurgery/Mectron might
negatively affect the EMC performances.
8.1.1
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions
PIEZOSURGERY® GP is intended for use in the electromagnetic environment specified below.
The customer or the user of PIEZOSURGERY® GP should ensure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Group 1
PIEZOSURGERY® GP only uses RF energy
for internal function. Therefore, its FR
emissions are very low and are not likely
to cause any interference with nearby
electronic equipment.
RF Emissions
CISPR 11
Class B
PIEZOSURGERY® GP is suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Emissions of fluctuations
voltage/flicker
IEC 61000-3-3
Compliant