USE
5.4
Important Information on Inserts
WARNING:
•
Diamond coated inserts
Diamond coated inserts are SINGLE USE.
The diamond coated inserts are intended
to be used on an individual patient
during a single surgical procedure and
then discarded.
The diamond coated inserts cannot
therefore be reprocessed since they
cannot be cleaned properly.
Bone and soil residues might remain
adhered to the diamond coating even
after cleaning and sterilization and enter
into the oral cavity of another patient
• When the titanium nitride coating is
visibly worn out, the insert must be
replaced. Use of an overly worn out
insert reduces its cutting efficiency.
• Do not activate the handpiece while
the insert is in contact with the treated
part, so that the electronic circuit can
recognize the best point of resonance
of the insert and allow its optimal
performance.
• Check the condition of wear of the
insert and that it is intact before and
during every use. In the event that a
drop in performance becomes apparent,
proceed to replace it.
•
Use original Piezosurgery/Mectron
inserts only. Use of non-original
inserts, in addition to voiding the
warranty, damages the threading of the
PIEZOSURGERY® GP handpiece, with
the risk of no longer being able to screw
the original inserts correctly during
subsequent use. Moreover, the device
settings are tested and guaranteed to
operate correctly only when original
Piezosurgery/Mectron inserts are used.
• Do not change the shape of the insert
in any way by either bending or filing it.
This could cause it to break.
•
Do not use an insert that has suffered
any type of deformation.
• Do not attempt to sharpen the insert
used.
• Always check that the threaded parts
of the insert and of the handpiece are
perfectly clean – see the Cleaning and
Sterilization Manual.
•
If excessive pressure is applied on the
insert, it can cause the insert to break
and possibly harm the patient.
• For information on how to correctly use
the inserts, refer to the annexed sheet
“Appropriate settings for the inserts
on the PIEZOSURGERY® GP” or the
illustrative leaflet of the Piezosurgery/
Mectron insert you’ve purchased.
• Before using PIEZOSURGERY® GP, make
sure you have prepared the operatory
site by having first moved away the soft
tissues, to avoid damaging them. It may
happen that, while cutting the bone,
accidental contact of certain parts of
the insert with the soft tissues inflicts
small traumas. To minimize this risk, use
specific protective instruments.