41
TROUBLESHOOTING
Problem
Possible Cause
Solution
No liquid flows out from the
insert during operation
The insert is the type without
a liquid passage
Use an insert type with a
liquid passage
The Water function is not
active
Activate the Water function
The insert is clogged
Unscrew the insert from the
handpiece and release the
water passage of the insert
by blowing compressed air
through it.
If the problem persists,
replace the insert with a new
one
The handpiece is clogged
Contact an Authorised
Mectron Service Centre
Poor performance
The insert is not correctly
tightened on the handpiece
Unscrew the insert and screw
it back in correctly using the
Mectron torque wrench (See
)
Insert broken, worn or
deformed
Replace the insert with a
new one
Table 6 –
Quick Troubleshooting