DISASSEMBLING PARTS FOR CLEANING AND STERILISATION
• In endodontic root canal therapy, never
operate the files outside of the root
canal to prevent them from breaking.
To prevent breakages, create a smooth
path with a manual endo file and plan
the most direct possible path of access
to minimise any bending of the insert.
Make delicate movements. Examine
the file often for signs of wear. If the file
breaks inside the canal, avoid contact
between the instrument and the broken
file to stop it from being pushed further
it. Do not apply pressure on the insert in
an axial direction.
5
DISASSEMBLING PARTS FOR CLEANING
AND STERILISATION
Before carrying out the cleaning procedures
described in
disconnect all accessories and components of
the COMPACT PIEZO P2K.
WARNING
: Switch the appliance off.
Always switch the appliance off using the
switch and disconnect the power supply
cable from the wall socket and machine
core unit before carrying out cleaning and
sterilisation tasks.
If present, remove the insert from the
handpiece using the torque wrench;
1
Disconnect the handpiece from the
relative cord;
2