Transmitter mag-flux M1

__    

Technical data

 

Reference conditions 

In conformity with IEC 770: 

temperature: 

T  = 20°C 

relative humidity: 

rH = 65%, 

air pressure: 

p  = 101,3 kPa 

Measuring tolerance 

See characteristic values of the corresponding sensor. 

Repeatability 

See characteristic values of the corresponding sensor. 

Influence of ambient temperature 

• 

For the pulse output: 

± 0,05 % per 10 K. 

• 

For the current output: 

± 0,1 % per 10 K. 

 

 

__   

Operating conditions

 

Installation conditions  

 

Caution 

Additional cable glands (not contained): 
The  operator  is  responsible  for  that  fact  that 
according  to  the  enclosure  and  ignition  enclo-
sure  certified  cable  glands  or  screws  are  used. 
The  kind  of  threads  is  stamped  on  the  rating 
plate. 
At  the  connection  between  sensor  and  trans-
mitter a metalized cable gland must be used for 
the screen. 
(See  chapter  „Connection  of  the  magnetic  cur-
rent and electrode line on page 9) 

 

Compact version 

For  the  compact  version  the  transmitter  housing  is  mounted  on 
the sensor. Therefore no cable  is  necessary between sensor and 
transmitter. 

Remote version 

The transmitter needs to be mounted separately from the sensor 
if: 

• 

the mounting area is difficult to access, 

• 

there is a lack of space, 

• 

medium and ambient temperatures are extremely high, 

• 

there is strong vibration. 

 

Fig 6 

Proper installation of cables at high humidity and wet conditions 

 

The 

mag-flux  M1

  transmitter  has  to  be  mounted  free  of  vibra-

tions! 

 

Caution: 

For  the  separate  version,  the  minimum  per-
missible  conductivity  of  the  medium  is  deter-
mined  by  the  distance  between  the  sensor  and 
the  transmitter.  The  maximum  cable  length  to 
ensure  accuracy  is  200 m.  For  the  cable  type 
see chapter „cable specification” on page 9. 

 

 

Fig 7 Cable length for remote version 

 
 
 

 

Important 

• 

The  electrode  cable  must  be  fixed.  If  the 
conductivity of the medium is low, cable move-
ments  may  change  the  capacity  considerably 
and thus disturb the measuring signal. 

• 

Do  not  lay  the  cables  close  to  electrical 
machines and switching elements. 

• 

Equipotential  bonding  must  be  ensured  be-
tween sensor and transmitter. 

 

 

Caution 

Do  not  connect  or  disconnect  the  field  coil  cable 
before the primary power of the meter has been 
disconnected! 

Environmental conditions 

Ambient temperature range 

- 20 °C to + 60 °C (-4°F to 140°F). 

Below 0 °C the readability of the LCD display will be limited. 

In  the  case  of  an  outdoor  installation,  the  device  must  be 
protected against direct solar irradiation with a weather shield 

Storage temperature 

- 25 °C to + 60 °C (-13 °F to 140 °F) 

Degree of protection 

IP67. 

 

Содержание mag-flux M1

Страница 1: ...a later time will not be honored II Warranty Your flowmeter was manufactured in accordance with the highest quality standards and was thoroughly tested prior to shipment However in the event any prob...

Страница 2: ...re to be used MECON GmbH accepts no liability for any loss or damage of any kind arising from improper operation of any product improper handling or use of any replacement part or from external electr...

Страница 3: ...trical devices device installation and process tech nology when replacing the transmitter elec tronics The highly integrated electronic com ponents of the device are ESD sensitive Caution The complete...

Страница 4: ...bes In order to be able to calculate the volume flow when using the mag flux flow probes correctly from the measured flow velocity the installation requirements must be kept regarding position and mou...

Страница 5: ...e operates the empty pipe detection Insulation coatings on the electrodes surface worse the empty pipe detection Data memory chip DSM The replaceable data memory chip DSM is an EEPROM device in DIL 8...

Страница 6: ...HART Volume flow or flow speed Using the HART protocol the current output has to be assigned to volume flow in the mode of 4 20mA Pulse Frequency output Pulse duration adjustable range is 0 1 2000 ms...

Страница 7: ...ds to be mounted separately from the sensor if the mounting area is difficult to access there is a lack of space medium and ambient temperatures are extremely high there is strong vibration Fig 6 Prop...

Страница 8: ...ss conditions Fluid temperature The data sheet rating plate of the connected transmitter is binding With directly mounted transmitter on the sensor compact version the heat entry from the process to t...

Страница 9: ...coil cable as shielded twisted pair In order to protect the cable from ex ternal interference the twisted pair wires are covered by an additional overall shield Cable length wire cross section Example...

Страница 10: ...spectively For terminal assignments see Electrical terminals on page 9 Caution Do not connect or disconnect the field coil cable before the primary power of the meter has been disconnected Please obse...

Страница 11: ...terface is connected via terminals 8 and 9 of the active current output The minimum load impedance is 250 Fig 19 Electrical connection for HART communication schematic diagram Display and operator int...

Страница 12: ...ry the display board inside the electronic compartment by loosing the three thread bolts 4 Unplug the green connector on the power supply board 5 Screw out the 3 thread bolts and after that extract th...

Отзывы: