Transmitter mag-flux M1
Connection of the magnetic current and electrode line
(remote version only)
Fig 14
Connection diagram for sensor cable of the
mag-flux M1
Terminal Function
5
Magnetic field current 1
6
Magnetic field current 2
7
Potential equilization / PE
22
Measuring ground
23
Electrode 1
24
Electrode 2
For cable specifications see chapter „Electrode cable and field coil
cable“ on page 9.
The outer shield is grounded by means of special EMC-compliant
cable glands at both ends of the cable, the inner shields are
connected to terminal 7 and 22 respectively.
For terminal assignments see „Electrical terminals“ on page 9.
Caution
Do not connect or disconnect the field coil cable
before the primary power of the meter has been
disconnected!
Please
observe
also
the
advices
in
chapter
„
Cable
specification
” on page 9
Connection of the sensor mag-flux A
The remote version of the sensor
mag-flux
A
has a terminal box
as shown in Figure 15.
Feed the electrode line through the left gland and the magnetic
current line through the right gland and connect the cables as
shown in Figure 15.
Fig 15
Electrical connections of the
mag-flux
A sensor
Connection of the sensor
mag-flux
F5
Feed the electrode line through the lower gland as shown in
Figure 16 and the magnetic current line through the upper gland
(see Figure 17) and connect the cables.
Fig 16
Electrical connections of the
mag-flux
F5 sensor (bottom)
Fig 17
Electrical connections of the
mag-flux
F5 sensor (top)