background image

 

Mecon GmbH   

Phone +49(0)241 41369 – 0 

Fax +49(0)241 41369 – 40 

Page 4 / 35 

Zieglerstraße 10 – 16   

www.mecon.de

 

[email protected]

 

Mecon 02/2010 

D-52078 Aachen   

 

 

 

Transmitter mag-flux M1 

 

Caution 

According  to  EN  50178:1997  all  electrical 
circuits  with    „protectiv  safety  isolation  with-
out  any  protection  against  contacts”  must 
observe the following maximum voltages: 

• 

Maximum AC voltage (V

eff

) 25 V 

• 

Maximum DC voltage 60 V 

It  is  strictly  forbidden  to  connect  FE  to 
any higher voltage! 

Zero point calibration 

In order to ensure that precise measurements are obtained, zero 
point calibration  is to be realized  the first time the device  is put 
into operation and before any regular operations are carried out. 
Zero point calibration is to be carried out using a fluid. 

The zero calibration procedure is as follows: 

• 

Install the sensor as described in the manual. 

• 

Check  to  ensure  that  the  sensor  is  completely  filled  with 
fluid and that there are no gas bubbles in the flow tubes. 

• 

Set  the  process  conditions  such  as  pressure,  temperature 
and density. 

• 

Close the cut-off device behind the sensor. 

• 

Operate  the  transmitter  in  accordance  with  the  instructions 
in chapter  „Zero point adjustment“ on page 11for the basic 
version  or  chapter  „Zero  point  calibration“  on  page  20  for 
the version with the control panel. 

• 

Make sure that  sufficient time  is  allowed for the electronics 
to warm up. 

• 

Allowing  fluid  to  flow  through  the  sensor  during  the  zero 
calibration  procedure  will  skew  the  zero  point  and  result  in 
false readings. 

Startup conditions 

The device is not subject to specific startup conditions. However, 
pressure surges should be avoided. 

Commisioning the 

mag-flux

 flow probes 

In order to be able to  calculate the volume flow  when using the 

mag-flux 

flow  probes  correctly  from  the  measured  flow  velocity, 

the installation requirements must be kept regarding position and 
mounting depth correctly. 

 

 

Transmitter  settings  must  be  made  as  specified  in 
chapter  „Operating  the 

mag-flux

  flow  probes  with 

the 

mag-flux  M1

”  on  page  6  to  ensure  the  correct 

operation!  

Particularly  for  existing  installations  after  replace-
ments or modifications e.g. tubing diameter. 

 
 

__   

Arbeitsweise und Systemaufbau

 

Measuring principle

 

It was back in 1832 that Faraday suggested utilizing the principle 
of  electrodynamic  induction  for  measuring  flow  velocities.    His 
experiments  in  the  Thames,  though  unsuccessful  due  to 
superimposed  polarization  effects,  are  nonetheless  regarded  as 
the  first  experiment  in  the  field  of  magnetic-inductive  flow 
measurement.    According  to  Faraday’s  law  of  electromagnetic 
induction,  an  electrical  field E  is  produced  in  a  conductive  liquid 
moving  through  a  magnetic  field B  at  a  velocity v  in  accordance 
with the vector product E = [v x B].   

 

Fig 2 Principle of the magnetic-inductive flow measurement 

 

Through  a  meter  tube  provided  with  an  insulating  lining  a  liquid 
flows  at  velocity v  and  a  flow  rate Q,  producing  a  measuring-
circuit voltage Um at the two electrodes at right angles to the di-
rection  of  flow. The size of this  measuring-circuit  voltage is pro-
portional to the mean flow velocity and the volume flow rate. 

System design 

The  meter  consists  of  a 

mag-flux  M1

  transmitter  and  a  sensor 

e.g.  mag-flux  series.    The  device  can  be  used  to  perform  mea-
surements  with  any  liquid,  conductive  media,  providing  that  the 
sensor’s material is suitable for the product being used. 

  
The 

mag-flux  M1

  transmitter  generates  the  inductive  current 

necessary  for  the  magnetic  field  and  preprocesses  the  induced 
voltage at the electrodes. 

Содержание mag-flux M1

Страница 1: ...a later time will not be honored II Warranty Your flowmeter was manufactured in accordance with the highest quality standards and was thoroughly tested prior to shipment However in the event any prob...

Страница 2: ...re to be used MECON GmbH accepts no liability for any loss or damage of any kind arising from improper operation of any product improper handling or use of any replacement part or from external electr...

Страница 3: ...trical devices device installation and process tech nology when replacing the transmitter elec tronics The highly integrated electronic com ponents of the device are ESD sensitive Caution The complete...

Страница 4: ...bes In order to be able to calculate the volume flow when using the mag flux flow probes correctly from the measured flow velocity the installation requirements must be kept regarding position and mou...

Страница 5: ...e operates the empty pipe detection Insulation coatings on the electrodes surface worse the empty pipe detection Data memory chip DSM The replaceable data memory chip DSM is an EEPROM device in DIL 8...

Страница 6: ...HART Volume flow or flow speed Using the HART protocol the current output has to be assigned to volume flow in the mode of 4 20mA Pulse Frequency output Pulse duration adjustable range is 0 1 2000 ms...

Страница 7: ...ds to be mounted separately from the sensor if the mounting area is difficult to access there is a lack of space medium and ambient temperatures are extremely high there is strong vibration Fig 6 Prop...

Страница 8: ...ss conditions Fluid temperature The data sheet rating plate of the connected transmitter is binding With directly mounted transmitter on the sensor compact version the heat entry from the process to t...

Страница 9: ...coil cable as shielded twisted pair In order to protect the cable from ex ternal interference the twisted pair wires are covered by an additional overall shield Cable length wire cross section Example...

Страница 10: ...spectively For terminal assignments see Electrical terminals on page 9 Caution Do not connect or disconnect the field coil cable before the primary power of the meter has been disconnected Please obse...

Страница 11: ...terface is connected via terminals 8 and 9 of the active current output The minimum load impedance is 250 Fig 19 Electrical connection for HART communication schematic diagram Display and operator int...

Страница 12: ...ry the display board inside the electronic compartment by loosing the three thread bolts 4 Unplug the green connector on the power supply board 5 Screw out the 3 thread bolts and after that extract th...

Отзывы: