Table 13. ACV AVG (60Hz Sine Wave Calibrating)
Range
Resolution
Accuracy
50Hz / 60Hz
50Hz - 1kHz
1kHz - 10kHz
10kHz - 20kHz
80mV
1
m
V
±(1.0%0 rdg + 8)
±(1.5% rdg + 8)
±(4.0% rdg + 8)
±(8.0% rdg + 8)
800mV
10
m
V
±(0.8%0 rdg + 8)
±(1.5% rdg + 8)
±(4.0% rdg + 8)
±(8.0% rdg + 8)
8V
0.1mv
±(0.8%0 rdg + 8)
±(1.0% rdg + 8)
±(3.0% rdg + 8)
±(5.0% rdg + 8)
80V
1mV
750V
10mV
50Hz - 400Hz : ±(1.0% rdg + 5)
Input impedance : 8mV - 800mV>1000M
V
, 8V - 1000V, 10M
V
, Parallel capacitance<100pF
Table 14. DCA
Range
Resolution
Accuracy
Note
80mA
1
m
A
800mA
10
m
A
± (0.5% rdg + 10)
Fuses : F 800mA / 250V F 10A / 250V
8A
0.1mA
Voltage drop :
i
800mV
10A
1mA
Table 15. ACA ( True RMS)
Range
Resolution
Accuracy
Note
80mA
1
m
A
400mA
10
m
A
50Hz - 2kHz
Fuses : F 800mA / 250V F 10A / 250V
8A
0.1mA
± (0.8% rdg + 10)
Voltage drop :
i
800mV
10A
1mA
Table 16. ACA AVG (60Hz Sine Wave Calibrating )
Range
Resolution
Accuracy
Note
80mV
1
m
A
400mV
10
m
A
50Hz - 500Hz
Fuses : F 800mA / 250V F 10A / 250V
8A
0.1mA
± (0.8% rdg + 10)
Voltage drop :
i
800mV
10A
1mA
Table 17. dBm
Function
Range
Accuracy
Resolution
Note
dBm
-80.00dBm--- + 80.00dBm
±1.0% rdg
0.01dBm
Table 18. Resistance(V)
Range
Resolution
Accuracy
Note
800
V
0.01
V
8K
V
0.1
V
Overload protection
80K
V
1
V
±(0.3% rdg+10)
250V RMS
800k
V
10
V
8M
V
100
V
80M
V
1k
V
±(2.5% rdg+8)
60M
V
-80M
V
: ±(3.5% rdg+10)
Table 23. Square wave output
OUT
Description
Voltage amplitude
3V approx.
Frequency
0.5Hz ~ 5000Hz
Duty cycle
1% ~ 99%
Table 24. Temperature
Temp
Accuracy
Resolution
Note
-50
0
C ~ 1372
0
C
±(2.5.% rdg+8)
1
0
C
type K themocouple
-58
0
F ~ 2502
0
F
1
0
F
Table 22. Capacitance
Range
Resolution
Accuracy
Note
500PF-1nF
1pF
±(5.0% rdg+10)
10nF
10pF
100nF
100pF
1
m
F
1nF
±(3.0% rdg+10)
Overload protection : 250V RMS
10
m
F
10nF
100
m
F
100nF
±(3.5% rdg+10)
Table 20. Frequenc (Hz)
Function
Range
Accuracy
Resolution
Note
999.99Hz
0.01Hz
9.9999kHz
0.1Hz
Overload protection : 250RMS
99.999kHz
±(0.05% rdg+5)
1Hz
Sensitivity : 0.7 RMS
Frequency
999.99kHz
10Hz
8.0000MHz
100Hz
10.0MHz
1kHz
100.0MHz
±(0.1% rdg+5)
10kHz
Plus adapter
1000.0MHz
100KHz
Table 21. Rotational speed (RPM)
Rotational
Range
Accuracy
Resolution
Note
speed
99999
±(0.1% rdg+5)
1RPM
Plus adapter
Table 19. Diode
Function
Range
Accuracy
Resolution
Note
Diode
8.0000V
±(0.3% rdg+5)
0.0001V
Diode positive Voltage drop
Overload protection: 250V RMS