APPENDIX 1   

A GUIDE TO THE SAFE USE OF MAINS POWERED TEST FRAMES

Always disconnect the machine from the mains power supply when not in use to avoid inadvertent actuation
of the machine by untrained personnel.

Protective Clothing

5

Eye protection should always be used in the form of an approved pair of safety spectacles. 
They should be replaced if they are scratched or damaged in any way. They should be kept clean and
clear to give maximum visibility to the user.

Extra bodily protection may be necessary if destructive testing or volatile failure of a test piece is likely. A risk
assessment should be carried out prior to using the test frame to ensure that all necessary safety measures have
been considered and carried out.

Machine Guarding

6

If, after the assessment, it is considered that machine guarding is needed, then contact your local 
supplier who, through Mecmesin, can arrange the supply of a suitable guard for the required level 
of protection.

Continued Safe Use

7

Once the machine is installed it should provide a reliable long term resource for universal testing. If 
however the machine fails, or appears to behave in an unusual manner, contact your local supplier for 
support. Do not continue to use the machine until it has been checked, and if necessary, repaired and 
returned to a safe working condition.

8

If the machine is damaged in use, advise your local supplier and have the machine repaired to a safe 
working condition. Do not use the machine until it is repaired.

9. The MultiTest 1 has not been designed for long duration tests at high speed and/or maximum load. A 

higher capacity stand (MultiTest 2.5) would be recommended for this type of application.

IF IN DOUBT CONSULT YOUR LOCAL SUPPLIER TO ENSURE CONTINUED SAFE USE

CAUTION: Hands must be kept clear of the crosshead and shroud opening at all times during
machine operation.

LIFTING INSTRUCTIONS: Before attempting to lift the MultiTest 1 turn off the power switch and unplug from
the mains. When lifting the MultiTest 1, support must be provided at the base, as well as at the back of the
shroud, with care being taken to avoid pressing against the adjustable limit switches.

Содержание MultiTest 1

Страница 1: ...It is essential that you familiarise yourself with the contents of these instructions before attempting to operate the MultiTest 1 MultiTest 1 Motorised Test Frame Operating Instructions ...

Страница 2: ......

Страница 3: ...ent immediately We strongly recommend that all the packaging and fixings are retained for any future transit requirements When using the MultiTest 1 please ensure that the ventilation holes situated on the back and bottom of the unit are not obstructed Also attach the four rubber feet to the MultiTest 1 with the screws provided Do not over tighten screws Before Use Appendix 1 gives guidance notes ...

Страница 4: ...munity Directives The MultiTest 1 has a single variable speed control for both upward and downward directions Speed can be set and controlled by the knob A in fig 1 on the front panel The white pointer on the speed control knob gives an indication of the speed setting Some test Standards place particular requirements upon speed accuracy reproducibility which may exceed the normal performance of th...

Страница 5: ...tion and then release The crosshead will move and the LED indicating direction of travel will illuminate When a limit switch position is reached the crosshead will begin to travel in the opposite direction at full speed When the position of the other limit switch is reached the motor will stop and the crosshead will come to rest During the cycle setting the Mode switch to Manual Control or selecti...

Страница 6: ...le damage would be to drive the test frame downwards until the loadcell or loadcell stud on AFG BFG instruments hits something solid This is a risk when a user is not yet familiar with operating a new test frame Consequently it is essential to set the limit switches B in fig 1 so that upwards downward travel of the crosshead is limited to prevent over travel and drive the gauge into the grips stan...

Страница 7: ...8in Operating modes Manual single cycle with fast return Reverse on alarm point Yes with AFG or AFTI cable Reverse on sample break Yes with AFG or AFTI cable ENVIRONMENTAL OPERATING CONDITIONS Temperature range 5ºC to 40ºC Humidity 92 5 STANDARD LOAD MEASUREMENT OPTIONS Force Gauge and dovetail bracket Yes Anvil plate Yes S beam Loadcell tension block module AFTI Yes SPECIAL OPTIONS Increased cros...

Страница 8: ...abel close to the mains inlet connector which advises which voltage is set for If the machine does not correspond to your supply inform your local Mecmesin supplier who will rectify the situation Connecting a mains powered test frame to the wrong supply will almost certainly cause extensive damage to the equipment The Mecmesin test frames must only ever be connected to a mains power installation t...

Страница 9: ...afe Use 7 Once the machine is installed it should provide a reliable long term resource for universal testing If however the machine fails or appears to behave in an unusual manner contact your local supplier for support Do not continue to use the machine until it has been checked and if necessary repaired and returned to a safe working condition 8 If the machine is damaged in use advise your loca...

Страница 10: ...arge M6 plain washers 5 Place all accessories listed to be despatched with the stand in the box with the test frame including the feet Ensure that the accessories and operators manual handbook are suitably wrapped for protection in transit 6 Place two of the four plastic foam spacers into the outside box so that they will support the ends of the inner box longest sides uppermost 7 Carefully place ...

Страница 11: ... outside the European Union EU Dovetail bracket fitted to the crosshead Appropriate mains cable Allen key for crosshead dovetail Four rubber feet four attachment screws and Allen key if necessary If you have any feedback regarding Mecmesin its products and services which you would like to share with us please contact us at f fe ee ed db ba ac ck k m me ec cm me es si in n c co om m page 9 ...

Страница 12: ... 1403 799979 f 44 0 1403 799975 Service Calibration t 44 0 1403 799920 f 44 0 1403 799925 i www mecmesin com e info mecmesin com Force and Torque Test Solutions DEALER STAMP 431 196 02 S I Instruments 256 South Rd Hilton South Australia 5033 Ph 08 8352 5511 info si instruments com www si instruments com ...

Отзывы: