A Guide to Safe Use of Mains Powered Test Systems
Mecmesin
5
preventing any movement of the crosshead or rotating platen. To re-set the E-stop button,
rotate it about 30 degrees clockwise.
2.4
Ventilation
To prevent overheating, ensure that all the air ventilation vents on the test system are not
obstructed. Where a console is fitted and if it is cooled by an internal fan, ensure the
console air vents are not obstructed.
Mecmesin test systems should only ever be installed in suitable
environmental conditions. The operating temperature and humidity should
be within the range given in the specification.
2.5
Electrical Supply and Earthing
Connecting a mains powered test frame to the wrong voltage supply will almost certainly
cause extensive damage to the equipment. Mecmesin test systems must only ever be
connected to a mains power installation that has a fully installed earthing system.
Connection of a mains powered test system to an electrical power outlet
without an earth connection is extremely dangerous and could lead to a risk
of electrocution. Consult your organisation’s qualified electrician if unsure.
The combined mains input socket and voltage selector can be found on the rear panel of
the test system (or on the right-hand side of twin-column force systems). Check that the
voltage selector is set correctly for your supply. The voltage that is selected is the one where
the arrows meet (the power inlet is inverted for some test systems):
▲
Selector set to 220-240 V
The voltage selection can be changed by switching off the power and removing the power
lead. The fuse cartridge can now be withdrawn. Rotate the fuse carrier and re-fit it so the
desired voltage marking is the correct way up.
See section 5.1 ‘Replacing fuses’ for how to remove the fuse carrier and replace a fuse.
Содержание MultiTest-0.5i
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 6: ...2 Mecmesin A Guide to Safe Use of Mains Powered Test Systems ...
Страница 16: ...12 Mecmesin A Guide to Safe Use of Mains Powered Test Systems ...
Страница 26: ...22 Mecmesin A Guide to Safe Use of Mains Powered Test Systems 7 1 3 431 DoC25 01 Vortex i to and xt derivative ...
Страница 27: ...A Guide to Safe Use of Mains Powered Test Systems Mecmesin 23 7 1 4 431 DoC26 01 Vortex dV ...
Страница 28: ...24 Mecmesin A Guide to Safe Use of Mains Powered Test Systems 7 1 5 431 DoC27 01 OmniTest 5 0 ...
Страница 29: ...A Guide to Safe Use of Mains Powered Test Systems Mecmesin 25 7 1 6 431 DoC28 01 OmniTest 10 to OmniTest 50 ...
Страница 30: ...26 Mecmesin A Guide to Safe Use of Mains Powered Test Systems 7 1 7 431 DoC30 01 Crimptest 1 0 ...
Страница 31: ...A Guide to Safe Use of Mains Powered Test Systems Mecmesin 27 7 1 8 431 DoC31 01 Helixa i and xt derivative ...
Страница 32: ......
Страница 33: ......