page 25
This function allows the average load reading to be
displayed. The average starts being calculated when the
START threshold (% of full-scale) is reached and stops
being calculated when the load passes through the
STOP threshold.
To set AVERAGE over TIME, press and hold the
MENU
key until page 1 of the main menu appears. Using the
UP
and
DOWN
keys, move the arrow cursor to AV/TIME
and press
ENTER
.
The display will show AV/TIME OFF and SET.
Press the
ENTER
key to change OFF to ON.
Press the
DOWN
key to move the arrow cursor to SET
and press the
ENTER
key.
The display will now show the START and STOP
thresholds and the values to which they are set (as a %
of full-scale).
Any load reading above the START threshold will be
averaged over time. Averaging stops when the load
reading passes through the STOP threshold.
A diamond cursor will indicate which value is selected.
Use the
UP
and
DOWN
keys to change the value, press
and hold to scroll values. When the correct value is
reached press the
ENTER
key to set START.
Repeat procedure for setting STOP.
The display will revert back to AV/TIME sub-menu 1.
AV/TIME
The maximum duration of AV/
TIME calculation is approx. 22
minutes.
AV/TIME sub-menu 1
(SET)
AV/TIME sub-menu 2
(PERCENTAGE START/STOP)
25N
100N
75N
50N
Start Threshold
x Peak
TIME
Stop Threshold
Average