page 16
The display shows OUT BAND and IN BAND. This menu
selects which values are to be considered.
OUT BAND Any value falling outside the set limits
LIMIT 1 and LIMIT 2.
IN BAND
Any value falling between the set limits
LIMIT 1 and LIMIT 2.
Use the
UP
and
DOWN
keys to move the cursor and
press
ENTER
key to select the desired feature.
The display shows PASS or FAIL. This menu sets the
OUT BAND criteria.
PASS
Values, which fall either OUT BAND (or
IN BAND, if selected), are a PASS and
will cause an audible beep, illuminate
an LED or both.
FAIL
Values, which fall either OUT BAND (or
IN BAND, if selected), are a FAIL and
will cause an audible beep, illuminate
an LED or both.
Use
UP
and
DOWN
keys to move the cursor and press
ENTER
key to select the desired feature.
The display shows BUZZER ON, CONTINUOUS and
PULSE. This menu selects the length of time that the
buzzer will sound, if AUDIBLE or BOTH has been
selected in
sub-menu 5.
CONTINUOUS The buzzer sounds at a pre-set alarm
value and stays on until the load falls
below that pre-set.
PULSE
The buzzer pulses for a fixed time of
one
second every time the load crosses
over
each of the presets.
Use the
UP
and
DOWN
keys to move the cursor and
press the
ENTER
key to select the desired feature.
The display will now return to the main menu page 1,
press
ESC
to return the main display.
ALARM sub-menu 5
(ALARM BAND)
ALARM sub-menu 6
(ALARM PASS/FAIL)
ALARM sub-menu 7
(ALARM BUZZER MODE)