2)    Operation 

 

a) It is recommended to complete the connection with the piezoelectric actuators before   

switching on the [POWER] switch.   

b) Make sure the bias voltage knob is set to zero if bias voltage is not needed. 

c) After switching on the

POWER

switch on the front panel, the device will be ready   

in function in a few seconds. The output voltage can be control with bias voltage   

knob or the input signal. 

 

3)    Over load protection circuit 

The built-in protector will be activated when the peak current is more than 5Amp or 

the  average  output  current  is  more  than  1  Amp.  When  the  built-in  protector  is 

activated, the

OVER L.

LED on the panel will turn on. Press the

RESET

switch 

to reset. Only reset the

RESET

switch after make sure there is no short circuit, 

overload and etc. 

 

4)    Power bandwidth 

For reference the power bandwidth of the MTAD3001, some examples of the driving 

conditions are summarized as follows. 

 

Load capacitance          Max. amplitude                Gain bandwidth (

-3

dB

 

3μF

       

     

     

100Vp-p     

           

DC

~   

2kHz

  *

 

1μF

           

     

 

100Vp-p     

           

DC

~   

6kHz

  *

 

(

  The  built-in  protector  activated  as  the  average  output  current  goes  above  the 

threshold.) 

           

 

5)    External control of built-in protector 

The built-in protector is designed to be access easily from customer control system. 

The  circuit  protector  can  be  controlled  remotely  from  an  external  circuit  by  TTL 

High signal level of control signals.The circuit is shown as follows. 

 

 