The Model PLA-600 are intended for use in the electromagnetic environment specified below.

Thecustomer or the user of the Model PLA-600 should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance level

Electromagnetic environment

-guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

± 6 kV contact ± 8

kV air

± 6 kV contact ± 8

kV air

Floors should be wood, concrete

orceramic tile. If floors are covered with

synthetic material, the relativehumidity

should be at least 30 %.

Electrostatic

transient / burst

IEC 61000-4-4

± 2 kV for power

supply lines ± 1

kV for

input/output lines

± 2 kV for power

supply lines

± 1 kV for

input/output lines

Mains power quality should be that of a

typical commercial or

hospitalenvironment.

Surge IEC
61000-4-5

± 1 kV differential

mode ± 2 kV

± 1 kV differential

mode± 2 kV

common mode common mode Mains

power quality should be that of a typical

commercial or hospitalenvironment.

Voltage dips, short

interruptions

andvoltage

variations on power

supply input lines

IEC 61000-4-11

< 5 % UT

(95 % dip in UT )

for 0.5 cycle

40 % UT

(60 % dip in UT )

for 5 cycles

70 % UT

(30 % dip in UT )

for 25 cycles

< 5 % UT

(>95 % dip in UT )

for 5 sec

< 5 % UT

(95 % dip in UT )

for 0.5 cycle

40 % UT

(60 % dip in UT )

for 5 cycles

70 % UT

(30 % dip in

UT ) for 25 cycles

< 5 % UT

(>95 % dip in UT )

for 5 sec

Mains power quality should be that of a

typical commercial or hospital

environment. If the user of themodels

PLA-600 product namerequires

continued operationduring power mains

interruptions, it is recommended that the

models PLA-600 be powered from an

uninterruptible power supply or abattery.

Power frequency

(50/60 Hz)

magnetic field IEC

61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should

be at levels characteristic of a typical

location in a typical commercial or

hospital environment.

NOTE: UT is the a. c. mains voltage prior to application of the test level.

Содержание PLA-600

Страница 1: ...logy Co Ltd No 11 District 9 Huayin Industrial Harbor No 618 West Kelin Road Chengdu Straits Science Tech Industrial Development Park Wenjiang District Chengdu China Customer Service 400 666 0366 Emai...

Страница 2: ...Electronic Technology Co Ltd with all rights reserved No part of this document may be used reproduced transmitted processed or recorded by any means or form electronic mechanical photographic or othe...

Страница 3: ...CH 3 7 3 1 CONNECTION GRAPH 3 3 2 WORKING PRINCIPLE 3 4 3 2 1 Principle 3 3 2 2 How It Works 3 4 3 2 3 Drawing of Principle 4 3 3 SYSTEM STRUCTURES AND CONFIGURATION INTRODUCTION 5 7 3 3 1 Introductio...

Страница 4: ...ortation 10 5 6 2 Storage 10 5 7 REGULATIONS ON SCRAPED EQUIPMENT 10 SECTION 6 SYSTEM SPECIFICATIONS 11 6 1 TECHNICAL SPECIFCIATION 11 6 1 1 Treatment Handle 11 6 1 2 Host Device 11 6 1 3 Power Output...

Страница 5: ...and turning on the unit To correct it please do as follows A Change the place of unit B Put the unit farther away from the influenced devices C Connect the unit to different electrical outlet or elect...

Страница 6: ...Attention Provides important matters requiring attention at first Section 1 Shows the purpose and organization of this manual Section 2 Describes the general configuration clinical indication and con...

Страница 7: ...re use the handle should be disinfected And there are bipolar and multi polar probes supplied 2 2 Clinical Indication and Contraindication 2 2 1 Indication Urology Surgeries BPH etc 2 2 2 Contraindica...

Страница 8: ...flows through human body to reach the electrode plate attached to patient thus making the energy through human body and surrounding tissues get much larger 3 2 2 How it Works To transmit the energy t...

Страница 9: ...nit Output Interface Chip Drive Display Sound Instruction Picture 1 The picture 1 shows the main structures of this device Signals are formed in the vibration unit and then output is realized in isola...

Страница 10: ...igns refers to the Mark of Flow Control Unit refers to the Mark of Protection Class ours is CF type refers to Footswitch Connection refers to Volume Adjustment at the Host Machine refers to Power Supp...

Страница 11: ...sion 01 3 3 2 Control Keys and Marking of the Host Device Shown with Graph 1 Manufacturer Logo 2 Four Pads 3 Socket for Footswitch 4 Socket for Treatment Handle 2 3 16 15 14 13 4 6 5 7 8 9 10 11 12 19...

Страница 12: ...output is closed operator can press the keys for change 13 Display of time After the device is started it will show the work time 14 Power Supply Display Light Green After power on the light is green...

Страница 13: ...to the socket at the front of host device and rotate it tightly 4 Plug the treatment handle into the corresponding socket of the device Before use make sure the treatment handle is cleaned and disinf...

Страница 14: ...il reaching the highest of 10 Gear After 10 the system will go back to 1 Gear 2 To reduce the gear position power output value press the key to reduce the voltage level of the corresponding mode Each...

Страница 15: ...the marking of Upwards Handle with Care Weather Proof 5 6 2 Storage This device should be stored in a well ventilated room without corrosive gas The height above ground and the distance from wall sho...

Страница 16: ...est Weight 10KG Height 19CM Width 39CM Length 45CM 6 1 5 Footswitch Cable length 4 5m Cleanness way anhydrous ethanol 6 2 System Classifications and Security Statement 6 2 1 System Classification As p...

Страница 17: ...curity Statement This unit meets the requirements of Term 19 1 19 2 and 19 3 of GB9706 1 2007 and GB9706 4 2009 Leakage electricity Grounding 200 A Leakage electricity Patients 20 A The unit can be us...

Страница 18: ...st Compliance Electromagnetic environment guidance 4 RF emissions CISPR 11 Group 1 The Model PLA 600 uses RF energy only for its internal function Therefore its RF emissions are very low and are not l...

Страница 19: ...ospitalenvironment Surge IEC 61000 4 5 1 kV differential mode 2 kV 1 kV differential mode 2 kV common mode common mode Mains power quality should be that of a typical commercial or hospitalenvironment...

Страница 20: ...es m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicin...

Страница 21: ...odel PLA 600 as recommended below according tothe maximum output power of the communications equipment Rated maximum output of transmitter W Separation distance according to frequency of transmitter m...