Mechan PLA-600 Скачать руководство пользователя страница 12

DOC. MEUM-131205700. Version 01

5.

Power Switch

6.

Model No. of the Unit: PLA-600

7.

Display Light for Treatment Probe Connection. Green when connected and Red when not connected.

8

Display Light for Footswitch Connection. Green when connected and Red when not connected.

9.

ABLATION Display to show the preset ABLATION output value in tow digits.

10.

A Key used to adjust the ABLATION output value: Every time the key is pressed, the value will be

increased or reduced by one level. (Attention: Only after the output is closed, operator can press the keys
for change).
11

PLACOAG Display to show the preset ABLATION output value in tow digits.

12.

A Key used to adjust the PLACOAG output value: Every time the key is pressed, the value will be

increased or reduced by one level. (Attention: Only after the output is closed, operator can press the keys
for change).
13.

Display of time: After the device is started, it will show the work time

14.

Power Supply Display Light (Green): After power on, the light is green.

15.

Impedance Display: real time display of the impedance value when the unit is started up.

16.

The Unit Label

17.

Nameplate of the Unit: to show the unit general information, e.g. unit no., factory date, power, fuse

electricity, etc.
18.

Fan Inlet.

19.

Fuse.

20.

Socket of Power Supply Cable: the cable is of three-line, so the electrical outlet should also be of

three-line with ground lead.
21.

Knob for Adjusting Device Volume: volume gets bigger when it’s turned in clockwise rotation or

vice versa.

22

Socket for Flow Control Unit.

23

GND

3.3.3

Identification of Working and Warning

1.

Working: There is discontinuous single sound of one time per second; working light is on; and the

treatment function is started up with output.
2.

Warning: The warning light glitters. The detected impedance is over high.

Содержание PLA-600

Страница 1: ...logy Co Ltd No 11 District 9 Huayin Industrial Harbor No 618 West Kelin Road Chengdu Straits Science Tech Industrial Development Park Wenjiang District Chengdu China Customer Service 400 666 0366 Emai...

Страница 2: ...Electronic Technology Co Ltd with all rights reserved No part of this document may be used reproduced transmitted processed or recorded by any means or form electronic mechanical photographic or othe...

Страница 3: ...CH 3 7 3 1 CONNECTION GRAPH 3 3 2 WORKING PRINCIPLE 3 4 3 2 1 Principle 3 3 2 2 How It Works 3 4 3 2 3 Drawing of Principle 4 3 3 SYSTEM STRUCTURES AND CONFIGURATION INTRODUCTION 5 7 3 3 1 Introductio...

Страница 4: ...ortation 10 5 6 2 Storage 10 5 7 REGULATIONS ON SCRAPED EQUIPMENT 10 SECTION 6 SYSTEM SPECIFICATIONS 11 6 1 TECHNICAL SPECIFCIATION 11 6 1 1 Treatment Handle 11 6 1 2 Host Device 11 6 1 3 Power Output...

Страница 5: ...and turning on the unit To correct it please do as follows A Change the place of unit B Put the unit farther away from the influenced devices C Connect the unit to different electrical outlet or elect...

Страница 6: ...Attention Provides important matters requiring attention at first Section 1 Shows the purpose and organization of this manual Section 2 Describes the general configuration clinical indication and con...

Страница 7: ...re use the handle should be disinfected And there are bipolar and multi polar probes supplied 2 2 Clinical Indication and Contraindication 2 2 1 Indication Urology Surgeries BPH etc 2 2 2 Contraindica...

Страница 8: ...flows through human body to reach the electrode plate attached to patient thus making the energy through human body and surrounding tissues get much larger 3 2 2 How it Works To transmit the energy t...

Страница 9: ...nit Output Interface Chip Drive Display Sound Instruction Picture 1 The picture 1 shows the main structures of this device Signals are formed in the vibration unit and then output is realized in isola...

Страница 10: ...igns refers to the Mark of Flow Control Unit refers to the Mark of Protection Class ours is CF type refers to Footswitch Connection refers to Volume Adjustment at the Host Machine refers to Power Supp...

Страница 11: ...sion 01 3 3 2 Control Keys and Marking of the Host Device Shown with Graph 1 Manufacturer Logo 2 Four Pads 3 Socket for Footswitch 4 Socket for Treatment Handle 2 3 16 15 14 13 4 6 5 7 8 9 10 11 12 19...

Страница 12: ...output is closed operator can press the keys for change 13 Display of time After the device is started it will show the work time 14 Power Supply Display Light Green After power on the light is green...

Страница 13: ...to the socket at the front of host device and rotate it tightly 4 Plug the treatment handle into the corresponding socket of the device Before use make sure the treatment handle is cleaned and disinf...

Страница 14: ...il reaching the highest of 10 Gear After 10 the system will go back to 1 Gear 2 To reduce the gear position power output value press the key to reduce the voltage level of the corresponding mode Each...

Страница 15: ...the marking of Upwards Handle with Care Weather Proof 5 6 2 Storage This device should be stored in a well ventilated room without corrosive gas The height above ground and the distance from wall sho...

Страница 16: ...est Weight 10KG Height 19CM Width 39CM Length 45CM 6 1 5 Footswitch Cable length 4 5m Cleanness way anhydrous ethanol 6 2 System Classifications and Security Statement 6 2 1 System Classification As p...

Страница 17: ...curity Statement This unit meets the requirements of Term 19 1 19 2 and 19 3 of GB9706 1 2007 and GB9706 4 2009 Leakage electricity Grounding 200 A Leakage electricity Patients 20 A The unit can be us...

Страница 18: ...st Compliance Electromagnetic environment guidance 4 RF emissions CISPR 11 Group 1 The Model PLA 600 uses RF energy only for its internal function Therefore its RF emissions are very low and are not l...

Страница 19: ...ospitalenvironment Surge IEC 61000 4 5 1 kV differential mode 2 kV 1 kV differential mode 2 kV common mode common mode Mains power quality should be that of a typical commercial or hospitalenvironment...

Страница 20: ...es m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicin...

Страница 21: ...odel PLA 600 as recommended below according tothe maximum output power of the communications equipment Rated maximum output of transmitter W Separation distance according to frequency of transmitter m...

Отзывы: