Push bu ons
Table 12. Push bu on specifica ons
Component
Label
Descrip on
Power bu on
POWER
W1 jumper installed (factory default) (
When device is off; press briefly to turn the device on
When device is on:
Press for ~1 second to shuts down the device; release
the bu on when the POWER LED blinks yellow
Press and hold ~4 seconds to force the device to power
off
W1 jumper removed (
Device will always be on if power is applied
Press for ~1 second to reboot the device; release the bu on
when the POWER LED blinks yellow
Func on bu on
FUNC
Dual func on:
Ejects, or safely unmounts, all removable media for safe
removal from the device (default)
This func on is disabled when a job that uses external
media is running
The MEDIA LED is steady green when all media is
unmounted but s ll inserted in the device
Starts or stops an acquisi on; requires configura on via the
web interface
Factory reset
bu on
CONFIG
RESET
Restores network se ngs to factory default values; refer to the
table for values.
Note 6. Refer to this
for the loca on of the W1 jumper.
Ground connector
Table 13. Ground connector specifica ons
Component
Label
Descrip on
Ground connector
#6-32
Connector port for the #6-32 ground screw
WebDAQ 316 User's Guide
page 27