Data Logging
Table 14. Data logging specifications
Parameter
Specification
Standalone power supply
USB power adapter
2.5 watt USB adapter with interchangeable plugs
(Includes plug for USA)
Memory card type
CompactFlash
Supplied memory card
512 MB CompactFlash card
Memory card host access
USB Mass Storage Device (MSD)
File systems supported
FAT16, FAT32
The device only creates 8.3 file names in the root subdirectory.
Log file format
binary
Logging rate
Min 1 second between entries, max 232 seconds, 1 second granularity
Data items logged
Timestamp, temperature, or raw reading from selected channels, state of DIO lines, CJC
sensor readings
Logging start methods
Configurable:
Start Logging on Power Up
– Logging begins 5 seconds after power on to allow
hardware to settle.
Start Logging on Button
– Device is idle on power on. Press and hold the button
until the LED turns on to begin logging. The first sample is acquired 1 second after the
LED turns on unless less than 5 seconds have elapsed since power on.
Start Logging at Specified Time
– Device is idle until the real-time clock
indicates the time is equal to or greater than the specified time, at which time the LED
turns on. The first sample is acquired 1 second after the LED turns on unless less than
5 seconds have elapsed since power on.
Logging stop methods
Stop on button press – To stop logging, press and hold the button until the LED turns off.
The device caches logged data in volatile memory prior to writing to memory card. When
logging, always use the button to stop logging and ensure data is written to memory card
prior to removing power.
Logging status indication
The LED operations when connected to the AC adapter power supply are different than
when connected to USB.
Logging modes:
Logging Off
mode: the LED is off (disabled).
Start Logging on Power Up
mode: the LED turns on, but blinks off momentarily
every time data is captured.
Start Logging on Button
mode: the LED is initially off. When the button is pressed
and held for approximately 1 second, the LED turns on and act the same as Start
Logging on Power Up mode.
Start Logging at Specified Time
mode: the LED turns off, with a momentary on
flash every second until the specified date/time is reached. At that time, the LED turns
on and acts the same as Start Logging on Power Up
mode.
Other indication:
To stop logging and store the remaining data to memory card, press and hold the button
until the LED turns off. It is then safe to remove the memory card.
If the memory card becomes full, the LED blinks rapidly (250 ms period).
If the memory card is removed while logging is in progress, the LED blinks rapidly
(250 ms period). Inserting a memory card stops the LED from blinking.
Real time clock
Table 15. Real time clock specifications
Parameter
Specification
Battery backup
CR-2032 lithium coin cell, replaceable
Accuracy
±1 minute per month