Measurement Specialties
KPSI
®
Digital Series User’s Manual
Page 26
www.Meas-Spec.com
XUT
Set Temperature
Units
aXUTu!
“u”=+0: Celsius
“u”=+1: Fahrenheit
:
aD0!
Example:
aXUT+1!
(Set units =Fahrenheit)
a0011
:
a (
Service Request
response)
Service Request
resp. returned in
001 seconds.
“1” is
number values
in“
D0
” resp:
au
“u” is +units selected.
XUU
Set User Units
aXUUso!
“s” is signed
user slope
.
“o” is signed
user offset
.
User output = (psi) * user
slope + user offset.
:
aD0!
Example:
aXUU+27.63+0!
(27.63 inches per psi)
a0012
:
a (
Service Request
response)
Service Request
resp. returned in
001 seconds.
“2” is
number values
in“
D0
” resp:
aso
“s” is user slope.
“o” is user offset.
NOTE:
A slope of 0 is invalid.
NOTE:
Be sure units of pressure
(“
XUP”
) are set to user units (9).